Standard Operations
5.4.4 Set the dilution factor
The Dilution Factor function is used to calculate the original
concentration of a sample that has been diluted by a known ratio.
For example, dilution factor can bring the analyzed concentration
within the test range.
1.
Press
Options
>
More…
>
Dilution Factor
.
The number entered at the dilution factor prompt will be
multiplied by the result to compensate for the adjustment.
For example, if the sample has been diluted by a factor of 2,
enter 2. The default setting of the dilution factor is turned off.
2.
Press
OK
to confirm. Press again
OK
.
3.
Press
Return
to return to the result display.
4.
Confirm ’Store present reading again?’ to save the current
measured value if necessary.
Note:
When a dilution is in effect, the dilution icon will appear on the
display.
Note:
If using undiluted samples, set the dilution factor off again.
5.4.5 Run a standard adjust
The Standard Adjust function allows the calibration curve for a
stored program to be adjusted based on analysis of a known
standard solution. The Accuracy Check section of written
procedures often suggests a standard solution concentration for
this purpose.
Содержание Portable Spectrophotometer DR 2800
Страница 2: ...2...
Страница 6: ...6 Table of Contents...
Страница 8: ...8 Specifications...
Страница 22: ...22 Installation...
Страница 116: ...116 Maintenance...
Страница 118: ...118 Troubleshooting...
Страница 120: ...120 Replacement Parts...
Страница 122: ...122 Contact Information...
Страница 124: ...124 Limited Warranty...