47
Standard Operations
Read a standard before setting Standard Adjust to On.
1.
Follow the entire procedure, using a known standard for the
sample.
2.
After reading the concentration, press
Options>More>Standard Adjust
.
3.
If Standard Adjust is set to
Off
, turn it
On
.
The Current Reading will show the concentration. The box on
the right will show the default standard value for the test, as
mentioned in the procedure.
4.
If the measurement used a standard concentration that is
different from the one displayed in the box, press the box on the
right to enter a different standard value and enter the new
value. Press
OK
to confirm.
5.
Press
Adjust
to enable the Standard Adjust. The Standard
Adjust icon will appear.
Note:
The adjustment must be within certain limits, which vary with
each program. The allowable percentage is shown after "Adjustment“.
Note:
When a Standard Adjust is in effect, the Standard Adjust icon
will appear on the display.
5.4.6 Set the chemical form
Some Stored Programs allow a variety of chemical forms to
be selected.
Press the unit (e.g. mg/L) or the chemical representation of the
evaluation form (e.g. Al
3+
). A list of available evaluation forms is
displayed. Select the required form by pressing the corresponding
entry in the list.
Note:
When the program is exited, the evaluation form reverts to the
standard setting.
An alternative way of changing the standard setting:
Содержание Portable Spectrophotometer DR 2800
Страница 2: ...2...
Страница 6: ...6 Table of Contents...
Страница 8: ...8 Specifications...
Страница 22: ...22 Installation...
Страница 116: ...116 Maintenance...
Страница 118: ...118 Troubleshooting...
Страница 120: ...120 Replacement Parts...
Страница 122: ...122 Contact Information...
Страница 124: ...124 Limited Warranty...