![Hach Portable Spectrophotometer DR 2800 Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/hach/portable-spectrophotometer-dr-2800/portable-spectrophotometer-dr-2800_user-manual_3160892049.webp)
49
Standard Operations
4.
Press
Options>More>Reagent Blank
.
5.
Press
On
to highlight the Reagent Blank function.
6.
The concentration shown on the key is the measured value of
the reagent blank. To use this value for more analyses of this
parameter, press
OK
.
7.
To enter a previously recorded reagent blank value, press the
key and use the alphanumeric keypad.
8.
Press
OK
.
Note:
The Reagent Blank function is deactivated when the measurement
program is left. To use the same blank value later for other tests using the
same reagent lot, enter the value per step
7
.
Note:
The results calculated using the reagent blank value must lie within
the limits of the measuring range of the test/method.
Note:
The reagent blank icon is shown in the result display (see arrow)
when the function is active.
5.4.8 Analysis of samples
1.
Press
Stored Programs
and select a program.
2.
Insert the blank cuvette/cell into the cuvette/sample cell holder.
3.
Press
Zero
.