The process of a calibration corresponds exactly to the process of a
normal measurement in the measurement operation. The only difference
is that instead of the current waste water sample the standard solution is
measured. The current output is not influenced by the calibration cycle.
The last measurement value before the calibration is recorded and is
only updated on completion of the calibration after a new measurement.
Automatic cleaning (approximately 15 minutes)
PHOSPHAX
Σ
sigma produces its cleaning solution by itself. To do this,
the digestion reagent A is fed into the cuvette. Then the thermal
conversion of the oxidant into sulfuric acid occurs. Now the entire
sample path to the end of the approximately 1.2 m long sample feed
hose outside the housing is filled with this hot acid. After 5 minutes of
reaction time, the used cleaning solution is rejected via the drain hose.
Safety panel
Within the digestion unit temperatures of over 100 °C are worked with,
under pressure and with a strongly acidic digestion agent. This requires
an increased safety effort.
For this reason, a safety panel is located in front of the entire analysis
part. On the internal side of the safety panel is the transparent safety
bolt, which is engaged in a special lock system behind the photometer
baseplate.
This lock only opens in the [+SERVICE] menu, after the device has
reached its default status (cuvette emptied, unpressurized and at
ambient temperature).
Dosing pumps
The four hose dosing pumps used are responsible for the transport of all
fluids.
They are stepper motor-controlled and variable in speed and direction of
rotation. The dosing hoses are routed with the help of pump cassettes
via the pump rollers. For the feeding of the reagents C and D, a two-
channel pump (two pump cassettes) is used, for the other pumps a one-
channel pump (one pump cassette) is used.
Wear on the pump rollers is indicated by strong groove formation, which
can lead to defective dosing. In this case, the exchange of the pump
rollers is recommended as a precaution.
Figure 9 Dosing pump head during roller exchange
N O T I C E
For all work inside the analysis part (dosing- and connecting hoses, pumps,
valves, digestion unit), this safety panel must first be removed and must be
positioned back before the measurement operation!
N O T I C E
For a long-term shutdown (over 48 hours) the
pump cassettes must be released in order to relieve the dosing hoses — but
because of the return flow risk the hoses should first be removed from the
screw caps (canister lids)! Please observe during the handling of the reagents the
safety notes under
section Exchange of the reagents
.
Pump cassette
Pump roller
Содержание PHOSPHAX sigma High Range
Страница 5: ...5 II 2 230 V AC 10 50 Hz 145 VA x 540 mm x 600 mm x 400 mm 22 kg...
Страница 7: ...7 12 2005 2002 96 Hach Lange Hach Lange...
Страница 8: ...8 2 SIGMATAX 2 PHOSPHAX sigma 50 kg 1785 1830 352 1200 527 527 300 0 5 C 40 C 420 1200 200 150 420...
Страница 9: ...9 1 SERVICE 2 3 5 6 3 4 4 3 CEE 10 A STO P 1 2 3 4 5 6 7 5 6...
Страница 10: ...10 4 1 2 m SIGMATAX 2 PG LZP 573 5 2 m 5 PG 6 9 mm 1 4 m T 2 5 A L 250 V...
Страница 11: ...11 6 SIGMATAX 2 SIGMATAX 2 1 2...
Страница 13: ...13 48 18 SERVICE PRE PUMPING 28 1 SERVICE 2 A C D 3 4 PRE PUMPING 5 6 PRE PUMPING 7 3 6 A C D 2...
Страница 15: ...15 15 PHOSPHAX sigma 1 2 m 5 100 C SERVICE C D 9 48...
Страница 16: ...16 3 A C D 6 6 1 2 a SERVICE 18 SERVICE 3 6 Hach Lange 10 30...
Страница 17: ...17 Q1 Q2 QA A A PHOSPHAX sigma 11 10 C 5 6 48 6 7 PHOSPHAX...
Страница 26: ...26 2 A BCF689 5 2 L 3 C BCF691 5 2 L 3 D BCF692 5 2 L 3 A BCZ822 250 mL D BCZ824 100 mL LCW 824 500 mL 6 12...
Страница 27: ...27 12 A D A D 5 3 1 2 D D 3 A D 2 x 1 4 A D 5 3 SERVICE 6 7 8 PRE PUMPING 6 12 1 2 3 4 SETTINGS STANDARD...
Страница 29: ...29 PHOSPHAX sigma 12 6 O A C D 3 3 18 24...
Страница 32: ...32...
Страница 114: ...Deutsch 114...
Страница 226: ...Fran ais 226...
Страница 282: ...Italiano 282...
Страница 370: ...Rom n 370...
Страница 452: ...T rk e 452...
Страница 456: ...456 2005 8 12 2002 96 EC Hach Lange Hach Lange...
Страница 463: ...463 100 C SERVICE C D 9 3 A C D 6 6 1 2 48 Pump cassette Pump roller...
Страница 464: ...464 SERVICE 466 SERVICE 3 6 Hach Lange 10 Q1 Q2 QA A A PHOSPHAX sigma 30 48...
Страница 465: ...465 11 STATUS 10 C 5 6 6 7 PHOSPHAX...
Страница 473: ...473 12 2 A BCF689 5 2 L 3 C BCF691 5 2 L 3 D BCF692 5 2 L 3 A BCZ822 250 mL D BCZ824 100 mL LCW824 500 mL 6 12...
Страница 476: ...476 PHOSPHAX sigma 12 6 O A C D 3 3 18 24...
Страница 477: ...477 Status STATUS SIGMATAX SIGMATAX STATUS SIGMATAX SIGMATAX SIGMATAX STATUS...
Страница 479: ...480...