Conductivity & TDS Operations
34
temperature compensation coefficient of 2% per °C. The coefficient is user
selectable in the setup mode.
Note: The meter automatically detects the cell constant. Calibration is automatically cleared if
the conductivity probe is changed to a probe of a different cell constant.
1.
Push
COND
or push
TDS
.
2.
Place probe in first standard solution. The meter will begin reading the
conductivity value.
3.
Push the arrow keys to scroll to the value of the conductivity or TDS standard
and push
ENTER.
The main display and the wait icon will flash.
4.
At any time during the calibration the user may exit the calibration mode by
pushing the
COND
,
TDS
or
SETUP
key. The meter will stop the current
calibration and keep the old calibration data.
5.
Push
CAL
. The main display value will begin flashing.
Note: To clear all previous calibration points, slopes and offset, push CAL then DELETE. To
update a previous calibration with a single point calibration, omit this step.
6.
Push
COND
or
TDS
to begin reading the conductivity or TDS value of any
solution.
7.
To have multiple conductivity or TDS calibration points, repeat the steps for up
to three calibration points.