pH Operations
22
2.
Push the arrow keys until the pH calibration point setup screen is displayed.
Push
ENTER
. The display will show the current number of calibration points.
Use the arrow keys to set the calibration points.
3.
Push
ENTER
to save changes and return to setup mode. Push
SETUP
to
return to setup mode without saving changes.
5.1.3 Set the pH calibration method
The meter allows the user to define which pH buffers are to be used, and in which
order. The default is a two
point calibration pH 7.00 for the first point and pH 4.00
for the second point.
To set the first point of a pH calibration method:
1.
Push
SETUP
then push
PH
to put in the pH setup mode.
2.
Push the arrow keys until the calibration method setup screen is displayed.