Operation
19
2.
Push the arrow keys until the clear all calibration points screen is displayed.
3.
Push
ENTER
.
4.
The meter will prompt if the user wants to clear all calibration points. The
default is No.
5.
Push the arrow keys to display Yes and then Push
ENTER
to clear.
6.
Push
SETUP
to return to setup mode without saving changes.
Note: This procedure DOES NOT clear calibration data for an ISFET probe.