Conductivity & TDS Operations
31
2.
Push the arrow keys until the conductivity probe constant screen is displayed.
Push
ENTER
.
6.2 Set the TDS options
6.2.1 Set the TDS conversion factor
TDS values are related to conductivity. The operator can calibrate using a TDS
standard or calibrate using conductivity standards then program the meter with a
conversion factor.
The conversion factor range is 0.40 to 1.00. The default is 0.50.
To set the TDS conversion factor:
1.
Push
SETUP
then push
TDS
.
2.
Push the arrow keys until the TDS conversion factor screen is displayed.