21
Section 5
pH Operations
5.1 Set the pH options
5.1.1 Set the pH resolution: 0.1 pH or 0.01 pH
The pH can be displayed in two resolutions: 0.1 pH or 0.01 pH. The stabilization
lock (
READY
icon) will occur faster at 0.1 pH resolution.
To select the resolution:
1.
Push
SETUP
then push
PH
to put the meter in the pH setup mode.
2.
Push the arrow keys until the pH resolution setup screen is displayed.
3.
Push
ENTER
. The display will show the pH resolution. Use the arrow keys to
set the pH resolution between 0.01 and 0.1.
4.
Push
ENTER
to save changes and return to setup mode. Push
SETUP
to
return to setup mode without saving changes.
5.
Push
ENTER
. The display will show the current number of calibration points.
Use the arrow keys to set calibration points between 2 and 3.
5.1.2 Set the 2 and 3 point pH calibration
The default calibration point is 2 point.
To select the number of calibration points:
1.
Push
SETUP
then push
PH
to put in the pH setup mode.