98
9.3 Troubleshooting the ORP Electrode
Table 17 Troubleshooting the ORP Electrode
Symptom
Probable Cause
Corrective Action
mV reading does
not correspond to
cited mV value for
standard.
Reference electrode/preamp
offset. The Ag/AgCl reference
half cell uses a 1.5 M KCl in
40% glycerol electrolyte
solution. This has a potential
of 32 +10 mV versus SCE. In
addition, the electrode
preamp can contribute a +35
mV offset. (See
Zero the preamp. See
Fouled sensor
Clean electrode in 1 M HCl. (See
Section 8.1.8 pH Electrode Cleaning)
ORP sensor is internally
shorted
Remove ORP sensor. Inspect it for wetness inside the glass stem. If wet,
replace. If dry, verify that two O-rings are being used, inspect them for
damage and replace if necessary. Completely dry both the inside of the
electrode socket and outside of the sensor assembly. Reinstall. (Refer to
Section 8.1.3 Sensor Replacement).
Reference has become
poisoned
Flush with large amounts of reference electrolyte. If this is not successful,
replace the entire electrode.
The mV readings
are erratic.
Air in the reference line
Verify that electrolyte is exiting the reference outlet. Wipe the end of the
electrode dry, and observe the reference outlet tube with the pump running
and the priming tee in the RUN position. A full drop of electrolyte should
become visible within 5 minutes (5 pumps). If a drop does not appear,
purge the system (see
Section 6.3 Purging the System) (also see
Symptom: No electrolyte flow).
Air in the sample line
If a pocket of air passes over or gets trapped at the end of the electrode, the
reading becomes momentarily erratic. This usually is remedied by
reorienting the electrode so that air bubbles do not block the sensor
(see
Section 5.5 Installing the pH/ORP Electrode).
ORP Sensor is broken.
Remove the ORP sensor from the electrode body and inspect it for wetness
within the glass stem (see
Section 8.1.3 Sensor Replacement) If wet,
replace sensor. If dry, verify that two O-rings are being used, inspect them
for damage and replace if necessary. Completely dry both the inside of the
electrode socket and outside of the sensor assembly. Reinstall. (Refer to
Section 8.1.3 Sensor Replacement).
There is no
electrolyte flow.
See
Содержание EC1000
Страница 2: ...2...
Страница 8: ...8...
Страница 12: ...12...
Страница 14: ...14...
Страница 16: ...16...
Страница 18: ...18 Figure 1 Controller Front Panel...
Страница 20: ...20 Figure 2 Pump Module...
Страница 24: ...24 Figure 6 pH Control Loop...
Страница 54: ...54...
Страница 56: ...56...
Страница 58: ...58 Figure 12 Controller Installation Panel Mount...
Страница 61: ...61 Figure 15 Installing the pH ORP Electrode Figure 16 Flow Thru Cell...
Страница 70: ...70...
Страница 73: ...73 6 4 Calibration See Section 2 3 Calibration Figure 21 Priming Tee Operation...
Страница 74: ...74...
Страница 80: ...80...
Страница 83: ...83 Figure 24 Sensor Replacement...
Страница 85: ...85 Figure 25 Hach One pH Process Electrode...
Страница 89: ...89 Figure 26 Pump Module Plumbing Diagram...
Страница 102: ...102...
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 108: ...108...
Страница 112: ...112...
Страница 116: ...116...
Страница 122: ...122...
Страница 123: ...123...
Страница 124: ......