60
Figure 14
Pump Installation
5.5
Installing the pH/ORP Electrode
Note: Improve stability in
high purity water by
grounding the solution
near the electrode.
Different mounting equipment is available for the three configurations for mounting the
electrode. The electrode can be configured as an in-line unit installed in a full pipeline (see
Figure 15) or on the end of a dip tube as a submersible unit in a tank or other vessel. The
electrode also can be placed in a flow-thru cell for applications of flow with very little
pressure (see Figure 16). Maximum pressure is 35 psi for in-line installation or 40 feet of
immersion depth for submersible installation. Do not deviate orientation more than
90 degrees from vertical for the three types of installation. The configuration used
determines whether the bushing or collar is threaded with the reference tube and cable
prior to installation. Maximum temperature for the electrode is 60 °C.
Содержание EC1000
Страница 2: ...2...
Страница 8: ...8...
Страница 12: ...12...
Страница 14: ...14...
Страница 16: ...16...
Страница 18: ...18 Figure 1 Controller Front Panel...
Страница 20: ...20 Figure 2 Pump Module...
Страница 24: ...24 Figure 6 pH Control Loop...
Страница 54: ...54...
Страница 56: ...56...
Страница 58: ...58 Figure 12 Controller Installation Panel Mount...
Страница 61: ...61 Figure 15 Installing the pH ORP Electrode Figure 16 Flow Thru Cell...
Страница 70: ...70...
Страница 73: ...73 6 4 Calibration See Section 2 3 Calibration Figure 21 Priming Tee Operation...
Страница 74: ...74...
Страница 80: ...80...
Страница 83: ...83 Figure 24 Sensor Replacement...
Страница 85: ...85 Figure 25 Hach One pH Process Electrode...
Страница 89: ...89 Figure 26 Pump Module Plumbing Diagram...
Страница 102: ...102...
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 108: ...108...
Страница 112: ...112...
Страница 116: ...116...
Страница 122: ...122...
Страница 123: ...123...
Страница 124: ......