Setup
(cont.)
INIT SET
C COMP (Inl)
MODE ( °C)
Type of temperature compensation (automatic, manual)
SET PNT ( °C)
Temperature setpoint for manual temperature compensation
(-5.0 to +105 °C)
SOL TC
Solution temperature compensation (- 0.1 pH to + 0.1 pH)
SERIAL (Inl)
BAUD (SEr)
Baud rate of serial I/O (300, 600, 1200, 2400, 4800, 9600)
PARITY (SEr)
Parity of serial I/O (odd, even, none)
DATA BIT (SEr)
Number of data bits of serial I/O (7/8)
STOP BIT (SEr)
Number of stop bits of serial I/O (7/8)
PRT DLY (SEr)
Enabled or Disabled
PRT INTR (SEr)
Print interval (00 to 99 MIN) (0 turns print interval off)
PRT LST (SEr)
Print a listing
OUTPUT 2
Output #2 Setup (Control, Record 2)
SIG AVG (Inl)
pH signal average value (00 to 10 SECS)
KEY LOCK
M UNLOCK (LoC)
Momentarily unlock the keyboard (30 minute time-out)
UNLOCK (LoC)
Unlock the keyboard
LOCK (LoC)
Lock the keyboard
REF SOLN
TMR RST (rEF)
Reset the reference solution timer
WARNINGS
DISP WRN
Display system warnings
CLR WARN
Clear system warnings
Table 20 Calibration Menu
Key
Option
Description
CAL
STD 1 (CAL)
AUTO STD (SD 1)
pH calibration point 1 with auto regulation of buffer value
MAN STD (SD 1)
pH calibration point 1 with manual entry of buffer value
STD 2 (CAL)
(Appears only after
STD 1 is completed)
AUTO STD (SD 2)
pH calibration point 2 with auto regulation of buffer value
MAN STD (SD 2)
pH calibration point 2 with manual entry of buffer value
GRAB SMP (CAL)
pH calibration using a grab sample
REVIEW (CAL)
OFFSET (CAL)
Display pH electrode offset calibration
SLOPE (CAL)
Display pH electrode slope calibration
STD 1 (CAL)
Display pH and mV value—use for cal point 1
STD 2 (CAL)
Display pH and mV value—use for cal point 2
TEMP (CAL)
TEMP 1 (CAL)
Temperature calibration point 1
TEMP 2 (CAL)
Temperature calibration point 2 (appears only after TEMP 1
is completed)
MV ZERO (CAL)
User input mv offset
Table 19 Setup Menu (continued)
Key
Option
Description
Содержание EC1000
Страница 2: ...2...
Страница 8: ...8...
Страница 12: ...12...
Страница 14: ...14...
Страница 16: ...16...
Страница 18: ...18 Figure 1 Controller Front Panel...
Страница 20: ...20 Figure 2 Pump Module...
Страница 24: ...24 Figure 6 pH Control Loop...
Страница 54: ...54...
Страница 56: ...56...
Страница 58: ...58 Figure 12 Controller Installation Panel Mount...
Страница 61: ...61 Figure 15 Installing the pH ORP Electrode Figure 16 Flow Thru Cell...
Страница 70: ...70...
Страница 73: ...73 6 4 Calibration See Section 2 3 Calibration Figure 21 Priming Tee Operation...
Страница 74: ...74...
Страница 80: ...80...
Страница 83: ...83 Figure 24 Sensor Replacement...
Страница 85: ...85 Figure 25 Hach One pH Process Electrode...
Страница 89: ...89 Figure 26 Pump Module Plumbing Diagram...
Страница 102: ...102...
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 108: ...108...
Страница 112: ...112...
Страница 116: ...116...
Страница 122: ...122...
Страница 123: ...123...
Страница 124: ......