47
Table 6 RS-232 Remote Commands
Code
Function
Response
GENERAL SETUP
CST
Do a cold start
Prints setup
SCD
Set calibration defaults
OK
SAV <=(0 to 10)>
Recall or set signal average
Value (on recall); or OK (on
set)
RCV
Recall current measurement values
Prints current values
KLO
Toggles keyboard lockout status and responds with the
new status
Locked or Unlocked
TCM<=(AUTO, MAN)>
Recall or set temperature compensation mode
Value (on recall); OK (on set)
TCS<=(-5 TO 105)>
Recall or set temperature compensation setpoint
Value (on recall); OK (on set)
RSR
Reference solution timer reset
OK
RECORDER
RMD <=(PH, MV, TEMP)>
Recall or set recorder mode
Value (on recall); OK (on set)
RMX <= -2 to 16 for pH,
±2000 for mV,
-5 to 105 for temperature>
Recall or set maximum value setpoint for current recorder
type
Value (on recall); OK (on set)
RMN <= -2 to16 for pH,
±
2000 for mV,
-5 to 105 for temperature>
Recall or set minimum value setpoint for current
recorder type
Value (on recall); OK (on set)
RTX <=(-2132 TO 2132)>
Move recorder trim maximum count by indicated amount
OK
RTN <=(-2132 to 2132)>
Move recorder trim minimum count by indicated amount
OK
CONTROLLER
CMD <=(pH, mV)>
Recall or set controller mode
Value (on recall); OK (on set)
CSP <=(-2 to 16 for pH,
±
2000 for mV)>
Recall or set controller setpoint
Value (on recall); OK (on set)
CDB <=(0 to 2 for pH,
0 to 100 for mV)>
Recall or set controller dead band
Value (on recall); OK (on set)
CPR <=(0 to 100)>
Recall or set controller preset
Value (on recall); OK (on set)
CGN <=(0 to 99.99)>
Recall or set controller gain
Value (on recall); OK (on set)
CRS <=(0.00 to 99.99)>
Recall or set controller reset value
Value (on recall); OK (on set)
CRT <=(0.00 to 99.99)>
Recall or set controller rate value
Value (on recall); OK (on set)
CTX <=(-2132 to 2132)>
Move controller trim maximum count by indicated amount
OK
CTN =(-2132 to 2132)
Move controller trim minimum count by indicated amount
OK
ALARMS
A1M <= (PH HIGH, PH LOW,
MV HIGH, MV LOW,
TEMP HIGH, TEMP LOW)
Recall or set alarm 1 mode
Value (on recall); OK (on set)
A1S <=(-2 to 16 for pH,
±
2000 for mV,
-5.0 to 105.0 for temp)>
Recall or set alarm 1 setpoint for current alarm type
Value (on recall); OK (on set)
A1H <=(0.01 to 2 for pH,
1 to 100 for mV,
0.1 to 2 for temp)>
Recall or set alarm hysteresis for current alarm type
Value (on recall); OK (on set)
A1D <=(0 to 30)>
Recall or set alarm 1 delay time
Value (on recall); OK (on set)
A2M <=(PH HIGH, PH LOW,
MV HIGH, MV LOW,
TEMP HIGH, TEMP/LOW)>
Recall or set alarm 2 mode
Value (on recall); OK (on set)
Содержание EC1000
Страница 2: ...2...
Страница 8: ...8...
Страница 12: ...12...
Страница 14: ...14...
Страница 16: ...16...
Страница 18: ...18 Figure 1 Controller Front Panel...
Страница 20: ...20 Figure 2 Pump Module...
Страница 24: ...24 Figure 6 pH Control Loop...
Страница 54: ...54...
Страница 56: ...56...
Страница 58: ...58 Figure 12 Controller Installation Panel Mount...
Страница 61: ...61 Figure 15 Installing the pH ORP Electrode Figure 16 Flow Thru Cell...
Страница 70: ...70...
Страница 73: ...73 6 4 Calibration See Section 2 3 Calibration Figure 21 Priming Tee Operation...
Страница 74: ...74...
Страница 80: ...80...
Страница 83: ...83 Figure 24 Sensor Replacement...
Страница 85: ...85 Figure 25 Hach One pH Process Electrode...
Страница 89: ...89 Figure 26 Pump Module Plumbing Diagram...
Страница 102: ...102...
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 108: ...108...
Страница 112: ...112...
Страница 116: ...116...
Страница 122: ...122...
Страница 123: ...123...
Страница 124: ......