39
SECTION 6 TROUBLESHOOTING
6.1 Adjusting the Calibration
If the instrument records consistently high or low values for the
accuracy check in Section 3.3, it may be necessary to adjust the
calibration as follows:
1. To enter the calibration mode, all channels must read END.
2. Simultaneously press and hold channel numbers 3 and 4. The
calibration screen will appear on the display. The screen will
show a channel number, the channel’s calibration value (Cal
Factor), and a current reading (Reading).
3. To change from one channel to the next, press the number for
the next desired channel.
Problem
Suggestion
Incomplete data set
Clear buffer.
Negative data points
Check temperature settings.
Incorrect data point (>1000)
Re-enter commands (see Table 4 on
page14). For example:
1. Select channel (i.e., S1).
2. Type “GA” to start data transfer.
Bubbles between tubing and cap
Re-seal with plastic tie-wrap.
Inconsistent stirring
Lift bottle and reposition on
instrument.
Records consistently low or high
values
See section 6.1 Adjusting the
Calibration.
Fan not working
Discontinue use. Contact the Hach
Service Department.
客服:400 688 9809
网址:www.mamots.com 电话:0431-85178766
Содержание BODTrak
Страница 1: ...400 688 9809 www mamots com 0431 85178766...
Страница 3: ...ii 400 688 9809 www mamots com 0431 85178766...
Страница 9: ...viii 400 688 9809 www mamots com 0431 85178766...
Страница 11: ...x 400 688 9809 www mamots com 0431 85178766...
Страница 13: ...2 400 688 9809 www mamots com 0431 85178766...
Страница 41: ...30 400 688 9809 www mamots com 0431 85178766...
Страница 48: ...37 400 688 9809 www mamots com 0431 85178766...
Страница 49: ...38 400 688 9809 www mamots com 0431 85178766...
Страница 51: ...38 400 688 9809 www mamots com 0431 85178766...
Страница 55: ...42 400 688 9809 www mamots com 0431 85178766...