6
•
The BODTrak method continuously removes carbon dioxide
(CO
2
) from the system so that the pressure difference
observed is proportional to the amount of oxygen used.
•
Bubbles formed by degassing can cause errors when samples
are warmed to the experimental temperature. The BODTrak
compensates for this occurrence by starting the test after the
equilibrium temperature has been reached.