SECTION 4 SPECIAL CONSIDERATIONS

4.1 Range and Volume Selection

Because the BODTrak instrument can measure various waste 
substances, a variety of total oxygen demand test results can be 
obtained. The instrument provides four direct-reading ranges 
(0-35, 0-70, 0-350, and 0-700 mg/L). Use Table 5 on page 20 to 
select the volume and range corresponding to the expected 
sample BOD. For example, when using a sample with an 
expected BOD of 35 mg/L or less, use a 420-mL volume.

For best results, analyze samples immediately after collection. If 
this is not possible, preserve samples at low temperature (4 °C) 
for no longer than 24 hours.

4.2 Sample Dilutions

If the sample’s BOD is unknown, you can generally assume that 
effluent is normally in the 0-70 mg/L range while influent is 
usually in the 0-700 mg/L range. 

If a sample does not contain sufficient nutrients for optimum 
bacteria growth, add the contents of one BOD Nutrient Buffer 
Pillow to each bottle. Do not add the BOD Nutrient Buffer Pillow 
if close simulation of original sample characteristics is required.

4.2.1 When Oxygen Demand Exceeds 700 mg/L
When the O

2

 demand of a sample exceeds 700 mg/L, dilute the 

sample with high-quality dilution water. Make the dilution water 
with distilled water that does not contain organic matter or traces 
of toxic substances such as chlorine, copper, and mercury. 

Demineralizers can release undetected organic matter that will 
create an objectionable oxygen demand. The most practical way 
to consistently produce water of low organic content is by 
distillation from alkaline permanganate. (For example, add 2 g 
KMnO

4

 and 4 g NaOH for every liter of water.)

After distillation, place 3 L of distilled water in a jug and bring 
the water temperature to 20 °C. Add the contents of one BOD 
Nutrient Buffer Pillow for 3 L to ensure sufficient nutrient 
concentration for diluted samples. Cap the jug and shake it 
vigorously for one minute to saturate the water with oxygen. Do 
not store the solution.

客服:400 688 9809 

网址:www.mamots.com                                                  电话:0431-85178766

Содержание BODTrak

Страница 1: ...400 688 9809 www mamots com 0431 85178766...

Страница 2: ...26197 18 BODTrak INSTRUMENT MANUAL Hach Company 1995 1998 All rights reserved Printed in the U S A ar lr 12 8 95 1ed Rev 5 7 98 400 688 9809 www mamots com 0431 85178766...

Страница 3: ...ii 400 688 9809 www mamots com 0431 85178766...

Страница 4: ...or certified compliance by Hach Company EN 50082 1 Immunity per 89 336 EEC EMC Supporting test records by Hach Company certified compliance by Hach Company Standards include IEC 801 2 and EN 61000 4 2...

Страница 5: ...y cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the in...

Страница 6: ...ng the Clock 10 2 4 RS232 Interface 11 2 4 1 Hardware Configuration 11 2 4 2 Commands 12 2 4 3 Downloading and Printing Test Results 16 SECTION 3 THE BOD TEST 19 3 1 BODTrak Test Procedure 19 3 2 Inte...

Страница 7: ...Toxic Materials 35 4 5 3 Seed Acclimatization 35 4 6 pH Effect 36 4 7 Supersaturation 36 SECTION 5 MAINTENANCE 37 5 1 Cleaning the BODTrak Instrument 37 5 1 1 Sample Bottles 37 5 1 2 Stir Bars and Se...

Страница 8: ...5 x 12 7 cm 13 x 12 x 5 inches External Power Supply Input 115 230 VAC 50 60 Hz Output 20 5 VRMS UL CSA and TUV approved Capacity Six 473 mL bottles Shipping weight 9 pounds Operating temperature 20...

Страница 9: ...viii 400 688 9809 www mamots com 0431 85178766...

Страница 10: ...hazard shall be used DANGER Indicates an imminently hazardous situation which if not avoided will result in death or serious injury WARNING Indicates a potentially hazardous situation that could resul...

Страница 11: ...x 400 688 9809 www mamots com 0431 85178766...

Страница 12: ...nviene leer las Fichas T cnicas de Seguridad y familiarizarse con los procedimientos de sugeridad ATTENTION La manipulation des chantillons chimiques talons et r actifs peut tre dangereuse Lire les fi...

Страница 13: ...2 400 688 9809 www mamots com 0431 85178766...

Страница 14: ...the BODTrak Instrument When removing the BODTrak instrument and accessories from the shipping containers inspect each item for damage that may have occurred during shipment The BODTrak shipping conta...

Страница 15: ...wer Supply switch located on the bottom of the Power Supply is set at 115V when shipped If 230 power is required move the switch to the 230V position 1 4 The BOD Test Biochemical Oxygen Demand BOD is...

Страница 16: ...ter sample Continuous stirring replenishes dissolved oxygen to the sample This makes the BODTrak results similar to occurrences found in a natural environment The dilution method supplies no additiona...

Страница 17: ...served is proportional to the amount of oxygen used Bubbles formed by degassing can cause errors when samples are warmed to the experimental temperature The BODTrak compensates for this occurrence by...

Страница 18: ...natural conditions 2 1 2 Pressure Sensor Function The BODTrak is sealed to prevent external atmospheric pressure changes from affecting the BOD reading Pressure sensors monitor air pressure within th...

Страница 19: ...ough six and are located in the center of the instrument control panel Scrolling Keys The display shows a graph of BOD values on the vertical axis and time in days on the horizontal axis During operat...

Страница 20: ...selection and exit the menu 5 Hold down the ON key A menu for selecting a BOD scale is displayed 6 Press the left or right key to select the desired scale This also initializes the current channel cl...

Страница 21: ...he channel stops The BOD curve is retained and can be accessed until another test is started on the channel If the test length is changed all channels are cleared CLR and data are erased Once ended a...

Страница 22: ...age 4 Connect the 9 pin D connector of a computer interface cable to the Serial I O port on the instrument Connect the opposite end of this straight through cable to the computer Serial I O port Com 1...

Страница 23: ...more Table 2 Wiring Cable from BODTrak to Computer BODTrak 9 pin Connector Socket Computer 9 pin Connector Socket Pin Signal Name Pin Signal Name 1 DCD 1 DCD 2 TXD 2 RXD 3 RXD 3 TXD 4 DSR 4 DTR 5 GND...

Страница 24: ...characters before the final carriage return and line feed General Conventions Commands preceded by S set conditions and are often followed by a single numeric parameter Commands preceded by G are que...

Страница 25: ...month to the 1 or 2 digit value xx SDxx Sets the current date to the 1 or 2 digit value xx Shxx Sets the current hour to the 1 or 2 digit value xx Smxx Sets the current minute to the 1 or 2 digit valu...

Страница 26: ...fter receiving the carriage return which ends the command Once started the data stream can not be interrupted S Sets the analyzer to demo mode where simulated BOD data are generated over several minut...

Страница 27: ...scroll bars Use function arrow and control keys for Windows 399 buffer lines NOTE Be sure to check inbound and outbound in Terminal Preferences to establish the PC BODTrak connection Then return to Se...

Страница 28: ...m is indicated by a message such as Test Run to Completion and a dollar symbol If small negative BOD values are observed at the beginning of a test see SECTION 6 TROUBLESHOOTING BOD Log for Ch 1 Statu...

Страница 29: ...Q 570 printer by pressing and holding the appropriate channel number for several seconds The instrument will send a copy of the graphical display as well as a truncated data stream 127 data points Thi...

Страница 30: ...o cuando manipule materiales muestras y desechos Use gafas de seguridad y guantes de caucho para evitar enfermedad o infecci n ATTENTION Suivre attentivement les techniques de laboratoire lors de la m...

Страница 31: ...to a BODTrak sample bottle see Table 5 See section 4 2 Sample Dilutions for more information on BOD range selection 3 Place a 3 8 cm 1 inch magnetic stir bar in each sample bottle Table 5 Selection of...

Страница 32: ...4 isoptional Ifsimulation of original samplecharacteristicsis required donotaddtheBOD NutrientBuffer 5 Apply Stopcock Grease to the seal lip of each bottle and to the top of each seal cup 6 Place a se...

Страница 33: ...niken beachten 7 Using the funnel add the contents of one Lithium Hydroxide Powder Pillow to each seal cup Do not allow lithium hydroxide particles to fall into the sample If this occurs discard the s...

Страница 34: ...channel until the stir bar is rotating properly 12 To select a test duration simultaneously press and hold the left and the right arrow keys until the time menu appears Press the CHANNEL 6 key to act...

Страница 35: ...ally 14 Press the ON key A menu for selecting the BOD range will be displayed 15 For 0 350 mg L range press the right key For 0 700 mg L press the key a second time For 0 35 mg L range press the left...

Страница 36: ...will change from RUN to END 17 Read the BOD results directly from the BODTrak display by pressing the key corresponding to each sample Note See section 2 4 3 Downloading and Printing TestResultsforin...

Страница 37: ...gen demand or nitrification These potential problems are discussed in detail in the next four sections 3 2 1 Bottle Leak A leak between the bottle cap and seal cup may cause readings similar to those...

Страница 38: ...ies of BOD tests using the same sample but different volumes or dilution ratios to select an appropriate BOD range Generally effluent normally is in the 0 70 mg L range while influent is in the 0 700...

Страница 39: ...ermedad o infecci n ATTENTION Suivre attentivement les techniques de laboratoire lors de la manipulation du mat riel des chantillons et des d chets Toujours porter des lunettes de s curit et des gants...

Страница 40: ...A 10 by volume of seed in solution will result see section 4 3 Sample Seeding 5 Follow the general procedure for the BOD test using the 0 350 mg L BOD range and a 5 day test period 6 Also perform a fu...

Страница 41: ...30 400 688 9809 www mamots com 0431 85178766...

Страница 42: ...OD Nutrient Buffer Pillow to each bottle Do not add the BOD Nutrient Buffer Pillow if close simulation of original sample characteristics is required 4 2 1 When Oxygen Demand Exceeds 700 mg L When the...

Страница 43: ...of Seed Certain types of BOD samples do not contain sufficient bacteria to oxidize the organic matter present in the sample Many industrial discharges are of this type Some sewage treatment plant eff...

Страница 44: ...mL Low seed concentrations are more critical than those that are too high They delay the start of oxidation and cause low BOD results Use the trial and error method to determine the optimum concentrat...

Страница 45: ...cate that the five day 20 C value can be obtained in two and one half days at 35 C Middlebrooks presents nomographs for converting BOD tests to temperatures other than 20 C 4 5 Industrial Wastes Indus...

Страница 46: ...concentrations of these and other materials see the Hach Water Analysis Handbook Dilute the sample with distilled water to eliminate the effect of these materials The correct BOD is obtained when two...

Страница 47: ...taining 10 of the waste material in question 4 Repeat steps 1 3 increasing the addition of waste material by 10 Stop the procedure when 100 waste material has been reached 4 6 pH Effect Low BOD test r...

Страница 48: ...37 400 688 9809 www mamots com 0431 85178766...

Страница 49: ...38 400 688 9809 www mamots com 0431 85178766...

Страница 50: ...its 3 Residues create a BOD so be sure to remove all detergents by rinsing the bottles several times with tap water and finally with distilled or demineralized water 5 1 2 Stir Bars and Seal Cups 1 Ri...

Страница 51: ...38 400 688 9809 www mamots com 0431 85178766...

Страница 52: ...current reading Reading 3 To change from one channel to the next press the number for the next desired channel Problem Suggestion Incomplete data set Clear buffer Negative data points Check temperatu...

Страница 53: ...eral times until the reading value approaches zero The value will fluctuate slightly from zero 7 To change the calibration factor press the left key to decrease the value and press the right key to in...

Страница 54: ...mpany customer service is an important part of every product we make With that in mind we have compiled the following information for your convenience GENERAL INFORMATION 400 688 9809 www mamots com 0...

Страница 55: ...42 400 688 9809 www mamots com 0431 85178766...

Страница 56: ...ask erlenmeyer 500 mL 505 49 Flask volumetric 1000 mL 547 53 Funnel analytical 65 mm 1083 97 Incubator undercounter Model 205 120V 26162 00 Incubator undercounter Model 205 240V 26162 02 Incubator Ref...

Страница 57: ...ck tube 75g 562 75 Lithium Hydroxide Powder Pillows 100 pkg 14163 69 Nitrification Inhibitor 35 g 2533 35 Polyseed Inoculum pkg 50 24712 00 Potassium Iodide Solution 100 g L 500 mL 12289 49 Potassium...

Страница 58: ...orado 80539 U S A By Telex 160840 Hach Loveland By FAX 970 669 2932 Hach Loveland Information Required Hach account number if available Billing address Shipping address Your name and phone number Purc...

Страница 59: ...s e de Namur 1 B 5150 Floriffoux Namur Belgium Telephone 32 8l 44 71 71 FAX 32 81 44 13 00 In Canada Hach Sales and Service Canada Ltd 1313 Border Street Unit 34 Winnipeg Manitoba R3H 0X4 Telephone 20...

Страница 60: ...eg Manitoba R3H 0X4 800 665 7635 Canada only Telephone 204 632 5598 FAX 204 694 5134 In Latin America the Caribbean the Far East the Indian Subcontinent Africa except Mediterranean Africa or Pacific B...

Страница 61: ...soon as possible always within two weeks of receipt The toll free number is 800 227 4224 RETURN OF ITEMS Authorization must be obtained from Hach Company before returning items for any reason For cred...

Страница 62: ...mmand summary 14 commands 12 communication software 11 computer 16 computer commands 11 12 computer interface 11 control keys 8 control panel 8 current time 14 cursor 9 curve 26 D data bits 11 data fl...

Страница 63: ...mmands 13 R range 24 31 RS232 11 S sample dilution 27 32 sample volume 20 scale 9 scrolling keys 8 seal cup 26 seed 27 29 Serial I O 11 Serial I O cable 11 software 11 specifications vii 3 standard sa...

Отзывы: