Transmitter 9125 – Conductivity / resistivity measurement
Page 39
S/MEASURE Menu
The PROBE menu allows to configurate the utilized
probe type, TEMP. COMP. the temperature
compensation, TDS the TDS measure coefficient
when it is actived.
PROBE
This menu allows to choose the utilized probe for the
conductivity measure.
TYPE :
- Induct.
- 2 elec.
choice of the electrode :
Inductive or 2 electrodes
Note :
Check both switches of the conductivity module are
correctly positioned :
- K : 2 electrodes
- I : inductive
K :
XXXX
Adjustment of the cell constant.
FREQ. : - Auto
- 8000
- 4000
- 2000
- 1000
- 500
- 250
- 125
- 62,5
Choice between an automatic
adjustment of the frequency
according to the measurement or
one of the pre-programmed
frequencies. (see Chapter
5 for
further details).
Only if elect. 2 type has been
selected.
TEMP. COMP.
TDS :
MEASURE
Select
PROBE
TYPE :
2 élect or inductive
K : 001.00
FREQ. :
PROBE
2 elect.
0.50
Auto