Transmitter 9125 – Conductivity / resistivity measurement
Page 49
S/RS485 Menu
This option requires the RS485 kit.
N° Monec
number
(0...32)
BAUD 300/600/1200/2400/4800/9600/19200
Transmission speed in bauds
PARIT.
- Without parity bit : No
- With odd parity bit : Odd
- With even parity bit : Even
BIT STOP
- 1 bit stop
- 2 bits stop
SWAP WORD
Allow to reverse the «
strong
weight », « light weight » size during
the manipulation of the real variable
(float type). Some equipment need
this reverse to read correctly the real
size data.
The communication protocol is MODBUS/JBUS.
The instrument may be equipped with a RS485 board
(optional) (see MODBUS 9100 manual).
N° : 4
BAUD : 9600
PARIT. : odd
BIT STOP : 1
SWAP WORD : NO
RS485
4