Transmitter 9125 – Conductivity / resistivity measurement
Page 30
Calibration Menu
Note :
Before any calibration launching, check the
parameters of the MEASURE menu (probe type,
temp. Comp.) Are correctly configured.
(See page 25).
This option allows to calibrate the conductivity
measure, the temperature measure and to display the
concentration measure (TDS).
Conductivity calibration
With the help of the « select » function touch, choose
COND. CALIB. option. Two options appears :
x
PROGRAMMING allows to choose the electric
conductivity calibration type :
¾
2 points
¾
1 point.
x
EXECUTION will allow to do the desired
calibration type.
Select to "Programming" and press enter.
Select the calibration type and press enter.
TDS CALIB
.
COND. CALIB.
TEMP.CALIB.
PARAMETERS
HISTORIC
CALIBRATION
Select
EXECUTION
PROGRAMMING
COND. CALIB.
Select
EXECUTION
PROGRAMMATION
COND. CALIB.
Select
PROGRAMMING
ETALONNAGE
MAINTENANCE
PROGRAMMING
SERVICE
MENU
Select
CALIBRATION