Transmitter 9125 – Conductivity / resistivity measurement
Page 34
TDS measure calibration
This option is available only if you have selected the
measurement TDS in the menu DISPLAY.
Select with the function key Choice
CALIB TDS
in the
menu
CALIBRATION
and press ENTER.
This calibration permits to adjust the coefficient K
TDS between the concentration value and
conductivity.
The measurement TDS is displayed as well as the
conductivity value as secondary measurement.
Proceed to the following steps :
x
Press the function key
OK
.
x
Change the displayed value and enter the sample
value (TDS).
x
The instrument indicates the result of the
coefficient calculation TDS and press ENTER.
Temperature calibration
With the help of the right function touch choice, select
the programming line and press ENTER.
The right function touch
allow to select 2
types of calibration, electric or process.
6.98
ppm
9.90 K
:
CALIBRATION
6.98
ppm
06.98 ppm
CALIBRATION
0