Orbisphere 510/511/512 Analyzers - Inputs/Outputs Menu
71 of 138
Operator Manual
ORBISPHERE
6.4 Analog Outputs
There are 3 analog outputs per channel. These outputs are configurable in terms of
function, content and behavior through the instrument menus. Analog outputs are used
to output a voltage or a current which is a function (e.g a linear characteristic) of a
measurement: AOut =
f
(M). Analog outputs can be typically connected to a PLC.
Knowing the function (
f
), the PLC can compute the value of the measurement.
Two types of instrument hardware are available:
• measurement board with current output (I = 0-20 mA or 4-20 mA)
• measurement board with voltage output (U = 0-5 V)
Fig 6-2:
Analog Outputs Menu
Содержание ORBISPHERE 510
Страница 1: ...Operator Manual ORBISPHERE 510 511 512 Revision G 05 11 2008 ...
Страница 2: ......
Страница 12: ...6 of 138 Table of Contents Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 43: ...Orbisphere 510 511 512 Analyzers View Menu 37 of 138 Operator Manual ORBISPHERE 3 View Menu Fig 3 1 View Menu ...
Страница 48: ...42 of 138 View Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 86: ...80 of 138 Inputs Outputs Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 110: ...104 of 138 Security Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 114: ...108 of 138 Global Configuration Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 116: ...110 of 138 Services Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE Fig 11 2 Services Menu Part 2 ...
Страница 134: ...128 of 138 Specifications Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 140: ...134 of 138 Glossary Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 143: ...Orbisphere 510 511 512 Analyzers User Notes 137 of 138 Operator Manual ORBISPHERE User Notes ...
Страница 144: ......