Orbisphere 510/511/512 Analyzers - Calibration Menu
61 of 138
Operator Manual
ORBISPHERE
5.3 TC Gas Sensor Calibration
5.3.1 Calibration of the Measured Gas
Start
Before initiating a calibration process, the calibration
parameters must be set by pressing on the
Modify
button. The last calibration parameters are memorized, so
this step can be ignored if the correct parameters are
already set.
Similarly, if only the calibration value has changed, then
this can be updated directly instead of pressing the
Modify
button.
Note:
When calibration is started, a calibration event is set, and
the analog output and relays for that channel are put on
hold (if selected below) to avoid unwanted alarms or
potential process problems. This resumes when
calibration is finished.
Modify Calibration Parameters
Gas Phase: Select
liquid
or
gas
(direct calibration
only)
Gas unit type:
Partial
,
fraction
or
dissolved
(dissolved is for calibration in a liquid only)
Gas unit: The list of available units depends on unit
type selected above.
Liquid: Select as appropriate.
Enter the gas concentration according to the value in
the calibration media.
⌧
Hold during calibration: On by default, this stops any
output from the instrument during the calibration
process to avoid sending invalid information to any
connected device.
⌧
Automatic calibration stop: If selected, when the
stability criteria is reached, the calibration process
stops automatically.
⌧
Interferences enabled: If selected, this takes into
account the influence of interferences during
calibration. By default the same interferences as
during measurement are selected.
Press OK to start calibration
Содержание ORBISPHERE 510
Страница 1: ...Operator Manual ORBISPHERE 510 511 512 Revision G 05 11 2008 ...
Страница 2: ......
Страница 12: ...6 of 138 Table of Contents Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 43: ...Orbisphere 510 511 512 Analyzers View Menu 37 of 138 Operator Manual ORBISPHERE 3 View Menu Fig 3 1 View Menu ...
Страница 48: ...42 of 138 View Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 86: ...80 of 138 Inputs Outputs Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 110: ...104 of 138 Security Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 114: ...108 of 138 Global Configuration Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 116: ...110 of 138 Services Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE Fig 11 2 Services Menu Part 2 ...
Страница 134: ...128 of 138 Specifications Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 140: ...134 of 138 Glossary Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 143: ...Orbisphere 510 511 512 Analyzers User Notes 137 of 138 Operator Manual ORBISPHERE User Notes ...
Страница 144: ......