![Hach Ultra ORBISPHERE 510 Скачать руководство пользователя страница 85](http://html.mh-extra.com/html/hach-ultra/orbisphere-510/orbisphere-510_operators-manual_577240085.webp)
Orbisphere 510/511/512 Analyzers - Inputs/Outputs Menu
79 of 138
Operator Manual
ORBISPHERE
These points should be set keeping in balance the following conditions (see
• The smaller is Range 2, the better is the analog signal resolution within the
expected range of measure.
• Size of Range 1 and 3 should be set to deliver an adequate level of resolution for
the measures falling out of the expected range of measure.
• In Range 0 the analog outputs only shows that measurement is below the MLL
value. Similarly in Range 4 the analog output only shows that measurement is
over the MHH value.
The formula to compute the measurement knowing the current or the voltage and the
resolution R is given in the following table:
6.5.3 “None” Analog Output
This is the default value.
Setting the analog output to "None" on a particular output means that analog output
value will always be zero and importantly ensures that no current is emitted on that
output, so reducing power consumption (especially useful in battery powered portables)
as well as reducing heat within the instrument.
Output
type
Tri-linear:
Range
Measurement M
Resolution R
4-20 mA
1: AOL > I > 4
M=MLL+(ML-MLL)
*
(I-4)/(AOL-4)
R=(ML-MLL)
*
20/((AOL-4)
*
1010)
2: AOH > I > AOL
M=ML+(MH-ML)
*
(I-AOL)/ (AOH-AOL) R=(MH-ML)
*
20/((AOH-AOL)
*
1010)
3: 20 > I > AOH
M=MH+(MHH-MH)
*
(I-AOH) / (20-AOH) R=(MHH-MH)
*
20/((20-AOH)
*
1010)
0-20 mA
1: AOL > I > 0
M=MLL+(ML-MLL)
*
I/AOL
R=(ML-MLL)
*
20/(AOL
*
1010)
2: AOH > I > AOL
M=ML+(MH-ML)
*
(I-AOL)/ (AOH-AOL) R=(MH-ML)
*
20/((AOH-AOL)
*
1010)
3: 20 > I > AOH
M=MH+(MHH-MH)
*
(I-AOH)/(20-AOH) R=(MHH-MH)
*
20/((20-AOH)
*
1010)
0-5 V
1: AOL > U > 0
M=MLL+(ML-MLL)
*
U/AOL
R=(ML-MLL)
*
5/(AOL
*
1010)
2: AOH > U > AOL
M=ML+(MH-ML)
*
(U-AOL)/ (AOH-AOL) R=(MH-ML)
*
5/((AOH-AOL)
*
1010)
3: 5 > U > AOH
M=MH+(MHH-MH)
*
(U-AOH) / (5-AOH) R=(MHH-MH)
*
5/((5-AOH)
*
1010)
Содержание ORBISPHERE 510
Страница 1: ...Operator Manual ORBISPHERE 510 511 512 Revision G 05 11 2008 ...
Страница 2: ......
Страница 12: ...6 of 138 Table of Contents Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 43: ...Orbisphere 510 511 512 Analyzers View Menu 37 of 138 Operator Manual ORBISPHERE 3 View Menu Fig 3 1 View Menu ...
Страница 48: ...42 of 138 View Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 86: ...80 of 138 Inputs Outputs Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 110: ...104 of 138 Security Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 114: ...108 of 138 Global Configuration Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 116: ...110 of 138 Services Menu Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE Fig 11 2 Services Menu Part 2 ...
Страница 134: ...128 of 138 Specifications Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 140: ...134 of 138 Glossary Orbisphere 510 511 512 Analyzers Operator Manual ORBISPHERE ...
Страница 143: ...Orbisphere 510 511 512 Analyzers User Notes 137 of 138 Operator Manual ORBISPHERE User Notes ...
Страница 144: ......