BIOMED SERVICE MANUAL Electrosurgical Generator G400 Workstation
Model 777000
Gyrus ACMI G400 Workstation
BIOMED SERVICE MANUAL
Page 17 of 57
Part Number: 177079-AA
Mains electrical safety checks
Perform the electrical safety tests detailed in section VIII and outlined in
IEC60601-1-1. Use a calibrated and approved safety tester designed for this
purpose. The test limits for the G400 generator are appropriate to Class1 type
BF. Use the Biomed Test Cable PN:BTK-G400 red, black and white wires to
access the patient applied parts of the generator.
Note that generator RF output power is available on demand when the Biomed
Test Cable is connected so observe all precautions cited in “PKRF output & RF
leakage checks general preamble”.
Record the test values and safety tester details on the form in section VIII.