7. If any questions or the generator does not meet the identified output
specification(s), call Gyrus ACMI Technical Services detailed on the contacts
page.
Video input output checks
This generator may be configured to superimpose information on to a Colour
Video Blanking & Sync (CVBS) video source presented at the VIDEO IN BNC.
A mixed CVBS signal is then available at the VIDEO OUT BNC. This
functionality is covered in the user manual for the generator. To test the video
function of the generator, connect the video output of the generator to a
monitor. With no electrodes attached, the generator should indicate Attach
Instrument. Hold down the keypad Mode switch until the display intensity
screen appears. Continue to press the Mode switch until Video OSD Select
appears. Set the option to ON then press Mode to return to Attach Instrument.
A blue screen should appear on the monitor with the text Attach Instrument
superimposed. If a CVBS video source is now connected to the Video In BNC
of the generator, the blue screen will change to the input video source with
Attach Instrument text superimposed.