BIOMED SERVICE MANUAL Electrosurgical Generator G400 Workstation
Model 777000
Gyrus ACMI G400 Workstation
BIOMED SERVICE MANUAL
Page 26 of 57
Part Number: 177079-AA
ERROR
CODE
REF
VFD TEXT
STRING
OSD TEXT
STRING
TYPE
DESCRIPTION
ACTION
600
07
INTERNAL
FAULT
!!!!
INTERNAL
FAULT !!!!
[F]
Aux board relay 5 fail Internal Fault. Return
for repair if error
persists
600
08
INTERNAL
FAULT
!!!!
INTERNAL
FAULT !!!!
[F]
Aux board relay 6 fail Internal Fault. Return
for repair if error
persists
600
09
INTERNAL
FAULT
!!!!
INTERNAL
FAULT !!!!
[F]
Aux board relay 7 fail Internal Fault. Return
for repair if error
persists
600
10
INTERNAL
FAULT
!!!!
INTERNAL
FAULT !!!!
[F]
Aux board relay 8 fail Internal Fault. Return
for repair if error
persists
600
11
INTERNAL
FAULT
!!!!
INTERNAL
FAULT !!!!
[F]
Aux board +5vPC_A
fail
Internal Fault. Return
for repair if error
persists
600
12
INTERNAL
FAULT
!!!!
INTERNAL
FAULT !!!!
[F]
Aux board +V_OSET
fail
Internal Fault. Return
for repair if error
persists
600
13
INTERNAL
FAULT
!!!!
INTERNAL
FAULT !!!!
[F]
Handswitching error
Internal Fault. Return
for repair if error
persists
GENERATOR DISASSEMBLY
The G400 generator has a Gyrus ACMI warranty seal that when broken, will
invalidate the warranty of the generator.
There are no customer serviceable components inside the generator case.
The generator casework is assembled with security screws that require special
tooling that is unavailable to third parties.