BIOMED SERVICE MANUAL Electrosurgical Generator G400 Workstation
Model 777000
Gyrus ACMI G400 Workstation
BIOMED SERVICE MANUAL
Page 21 of 57
Part Number: 177079-AA
ERROR
CODE
REF
VFD TEXT
STRING
OSD TEXT
STRING
TYPE
DESCRIPTION
ACTION
200
27
SYSTEM
RESET
!!!!
INTERNAL
FAULT !!!!
[F]
PlasmaBlend
Transformer
feedback failure
Internal Fault. Return
for repair if error
persists
200
28
INTERNAL
FAILURE /
WARMING
UP
!!!!
INTERNAL
FAULT!!!
WARMING
UP
[SOF
T]
[F]
G400 Workstation
too cold, dependent
on temperature.
Allow to reach room
temperature.
None
200
29 - 30
-
-
-
Reserved
200
31
INTERNAL
FAILURE
!!!!
INTERNAL
FAULT !!!!
[F]
CURRLIM signal
error (DAC output)
Internal Fault. Return
for repair if error
persists
200
32
INTERNAL
FAILURE
!!!!
INTERNAL
FAULT !!!!
[F]
VOLTLIM signal
error (comparator
input )
Internal Fault. Return
for repair if error
persists
200
33
INTERNAL
FAILURE
!!!!
INTERNAL
FAULT !!!!
[F]
BUSVOLTS signal
error (analogue input
)
Internal Fault. Return
for repair if error
persists
200
34
INTERNAL
FAILURE
!!!!
INTERNAL
FAULT !!!!
[F]
Incorrect PK RF
board installed
Internal Fault. Return
for repair if error
persists
200
35
INTERNAL
FAILURE
!!!!
INTERNAL
FAULT !!!!
[F]
RF relay 1 (socket)
not operating
Internal Fault. Return
for repair if error
persists
200
36
INTERNAL
FAILURE
!!!!
INTERNAL
FAULT !!!!
[F]
RF relay 2 (poles)
not operating
Internal Fault. Return
for repair if error
persists
200
37-41
-
-
-
Reserved
200
42
INTERNAL
FAILURE
!!!!
INTERNAL
FAULT !!!!
[F]
CPU POST
Thermistor failure
Internal Fault. Return
for repair if error
persists
200
43
INTERNAL
FAILURE
!!!!
INTERNAL
FAULT !!!!
[F]
CPU POST failure
Internal Fault. Return
for repair if error
persists
200
44
RAM
!!!!
INTERNAL
FAULT !!!!
[R]
RAM test failure
[POST check]
Internal Fault. Return
for repair if error
persists
200
45
INTERNAL
FAILURE
!!!!
INTERNAL
FAULT !!!!
[F]
Crystal failure [POST
check]
Internal Fault. Return
for repair if error
persists
200
46
-
-
-
NOT USED
200
47
INTERNAL
FAILURE
!!!!
INTERNAL
FAULT !!!!
[F]
Analogue reference
failure [POST check]
Internal Fault. Return
for repair if error
persists
200
48
INTERNAL
FAILURE
!!!!
INTERNAL
FAULT !!!!
[F]
CLAMP_SET failure
[POST check]
Internal Fault. Return
for repair if error
persists
200
49
INTERNAL
FAILURE
!!!!
INTERNAL
FAULT !!!!
[F]
BOOST_SET failure
[POST check]
Internal Fault. Return
for repair if error
persists
200
50
INTERNAL
FAILURE
!!!!
INTERNAL
FAULT !!!!
[F]
CURRLIM_SET
failure [POST check]
Internal Fault. Return
for repair if error
persists