BIOMED SERVICE MANUAL Electrosurgical Generator G400 Workstation
Model 777000
Gyrus ACMI G400 Workstation
BIOMED SERVICE MANUAL
Page 25 of 57
Part Number: 177079-AA
ERROR
CODE
REF
VFD TEXT
STRING
OSD TEXT
STRING
TYPE
DESCRIPTION
ACTION
400
17
BUTTON
STUCK
BUTTON
STUCK!
[R]
Front panel : Blue
UP button stuck
Don’t depress buttons
during power up.
Return for repair if
error persists
400
18
BUTTON
STUCK
BUTTON
STUCK!
[R]
Front panel : Blue
DOWN button stuck
Don’t depress buttons
during power up.
Return for repair if
error persists
400
19
BUTTON
STUCK
BUTTON
STUCK!
[R]
Front panel : MODE
button stuck
Don’t depress buttons
during power up.
Return for repair if
error persists
400
20
BUTTON
STUCK
BUTTON
STUCK!
[R]
Front panel :
STANDBY button
stuck
Don’t depress buttons
during power up.
Return for repair if
error persists
400
21
-
-
-
Reserved
400
22
HANDSWIT
CH STUCK
HDSWT
STUCK!
[AF]
Handswitch Yellow
button stuck
Don’t depress buttons
during power up.
Return for repair if
error persists
400
23
HANDSWIT
CH STUCK
HDSWT
STUCK!
[AF]
Handswitch Blue
button stuck
Don’t depress buttons
during power up.
Return for repair if
error persists
400
24
HANDSWIT
CH STUCK
HDSWT
STUCK!
[AF]
Handswitch Mode
button stuck
Don’t depress buttons
during power up.
Return for repair if
error persists
400
30
-
-
-
Reserved
500
10
-
-
-
Reserved
600
01
SYSTEM
RESET
!!!!
INTERNAL
FAULT !!!!
[F]
POST Aux board
ADC check. Raised if
the Aux board ADC
does not respond in
time
Internal Fault. Return
for repair if error
persists
600
02
INTERNAL
FAULT
!!!!
INTERNAL
FAULT !!!!
[F]
Aux board relay
checks
Internal Fault. Return
for repair if error
persists
600
03
INTERNAL
FAULT
!!!!
INTERNAL
FAULT !!!!
[F]
Aux board relay 1 fail Internal Fault. Return
for repair if error
persists
600
04
INTERNAL
FAULT
!!!!
INTERNAL
FAULT !!!!
[F]
Aux board relay 2 fail Internal Fault. Return
for repair if error
persists
600
05
INTERNAL
FAULT
!!!!
INTERNAL
FAULT !!!!
[F]
Aux board relay 3 fail Internal Fault. Return
for repair if error
persists
600
06
INTERNAL
FAULT
!!!!
INTERNAL
FAULT !!!!
[F]
Aux board relay 4 fail Internal Fault. Return
for repair if error
persists