SPECTRUM ANALYZER
259
Stop: 500Hz ~ 750MHz
Span: 1kHz ~ 500MHz
Peak to center
Pressing this button will set the
frequency location of the spectrum
peak as the new center frequency of
the Spectrum Analyzer.
Configuring the Bandwidth
Background
The bandwidth menu gives the possibility of
configuring the resolution bandwidth as well as
the type of window used for the spectral analysis.
Panel Operation
1.
Press the
Option
key, and
the
Spectrum Analyzer
button to enter the
Spectrum Analyzer menu.
Option
2.
Press the
Bandwidth
button to enter
the bandwidth menu.
3.
The resolution bandwidth
can be
set automatically according to a
configurable ratio defined between
the span and the frequency
resolution. To choose that option,
set the RBW
Mode
button to
Auto
,
press on the
Span: RBW
button and
tune the ratio using the variable
knob.
Range
5,000:1 ~ 1,000:1
4.
Alternatively, set the
RBW. Mode
button to
Manual
to manually
configure the frequency resolution.
All manuals and user guides at all-guides.com