MDO-2000E series User Manual
168
I
2
C Bus Trigger Settings
The I
2
C bus trigger conditions can be set at any time after the bus
settings has been set to I
2
C.
Panel Operation
1.
Set the Bus to I
2
C in the bus menu.
Page 124
2.
Press the
Trigger Menu
key.
Menu
3.
Press
Type
from the bottom menu.
4.
Press
Others
from the side
menu and select
Bus
.
The Trigger on settings will be reflected on the
Trigger Configuration icon.
From left: Bus trigger, Trigger source
5.
Press
Trigger On
and select the
triggering condition for the
selected bus.
Trigger On
Start, Repeat Start, Stop, Missing
Ack, Address, Data, Address/Data
Trigger On – Data
If Data or Address/Data was configured for the
Trigger On setting, then the number of bytes,
data and addressing mode (I
2
C) can be
configured.
6.
Press
Data
from the bottom menu.