ADVANCED CONFIGURATION
173
The Trigger on settings will be reflected on the
Trigger Configuration icon.
From left: Bus trigger, Trigger source
5.
Press
Trigger On
and select the
triggering condition for the
selected bus.
Trigger On
Start of Frame, Type of Frame,
Identifier, Data, Id & Data, End of
Frame, Missing Ack, Bit Stuffing Err
Trigger On –Type
of Frame
6.
If
Type of Frame
was configured for the Trigger
On setting, then the type of frame can be
configured from the side menu.
Type
Data Frame, Remote Frame, Error
Frame, Overload Frame
Trigger On –
Identifier
7.
If
Identifier/Id & Data
was configured for the
Trigger On setting, select the format from the
side menu.
Format
Standard, Extended
8.
Press
Identifier
from the side menu
to set the identifier data.
To edit the identifier, use the
Variable
knob to highlight a binary
or hex digit and press
Select
. Use
the
Variable
knob to choose a value
for the digit and press
Select
to
confirm.
VARIABLE
Binary
0,1,X (don’t care)
Hex
0~F, X (don’t care)