ARBITRARY WAVE GENERATOR
217
4.
Press
Select
from the side menu
and choose GEN1 or GEN2.
Turn the Output On for the Selected Channel
Background
The output for each generator channel can be
turned on or off independently.
Panel Operation
1.
Press
Output
to toggle the selected
channel on or off.
Setting the Load Impedance
Background
The load impedance can be independently set for
each generator channel.
Panel Operation
1.
Press
Load
to toggle the impedance
between 50Ω and High Z.
Setting the Phase
Background
The output phase can only be set for the GEN1
output. GEN2 is always set to an output phase of
0°.
Panel Operation
1.
Press
Phase
and use the variable
knob to set the phase.
Phase
-180° ~ 180°
Reset Phase
2.
The phase can be reset by pressing
S_Phase
.
All manuals and user guides at all-guides.com