MDO-2000E series User Manual
146
Trigger Parameter Overview
All the following parameters are common for all
the trigger types unless stated otherwise.
Trigger Source
CH1 ~ 4
Channel 1 ~ 4 input signals
Except for: Logic
EXT *
External trigger input
signal
Except for: Video, Pulse
Runt, Rise & Fall, Bus and
Logic
*Only for 2 input models.
EXT TRIG
AC Line
AC mains signal
Except for: Video, Pulse Runt, Rise & Fall,
Bus and Logic
D0 ~ D15
Digital input channels
Except for: Video, Pulse Runt, Rise and
Fall
Alternate
Alternate between channel sources for
the trigger source.
EXT Probe
For EXT trigger source only. Set the
probe as either current or voltage.
Attenuation
For EXT trigger source only.
Attenuates the EXT trigger probe by
an adjustable value.
Range
0.001X ~ 1000X 1-2-5 steps
Source Bus
UART
UART bus
I
2
C
Inter-Integrated Circuit
SPI
Serial Peripheral Bus
Parallel
Parallel bus
CAN
Controller Area Network bus
All manuals and user guides at all-guides.com