LCR-6000 Series User Manual
57
θd
Phase angle
R
Resistance (=Rs)
X
Reactance
G
Conductance
B
Sustenance
Y
Absolute value of admittance
5.1.6 Measurement Delay [DELAY]
This setting decides how long the meter will wait to
commence a measurement after the trigger signal is
received. Settable range: 0ms to 60s.
5.1.7 AUTO LEVEL CONTROL [ALC]
The automatic level control (ALC) feature adjusts the
voltage across the DUT or the current through the DUT
to match the voltage/current level setting. Using this
feature, you can try to ensure a constant signal level
(voltage or current) is applied to the DUT.
NOTE
In situations when the actual measuring Vac or
Iac goes beyond the extent that the ALC can
regulate, a warning message, “Failed! ALC
can’t regulate!”, will be shown at the bottom of
the screen to indicate to the operator that the
ALC can no longer regulate Vac/Iac.
An asterisk will be shown beside the LEVEL V
or A unit when the ALC is turned on.
Soft-key
Function
OFF
Turn off the Auto Level Control
Function
ON
Turn on the Auto Level Control
Function
5.1.8 Nominal Value [NOMINAL]
The Nominal Value parameter becomes displayed and
settable only when MON1 and MON2 are set in either Δ
or Δ% modes. The Nominal Value and the measured
main parameter will then be used to perform the Δ or
Δ% calculations.