LCR-6000 Series User Manual
103
Query Response <0-8><NL>
Example
SEND> FUNC:DCR:RANG?<NL>
RET>
0<NL>
11.7.5 FUNCtion:RANGe:AUTO
The FUNCtion:RANGe:AUTO command sets the auto range to ON or OFF.
Command
Syntax
FUNC:RANGe:AUTO {off(hold),on(auto),NOMinal}
Parameter
Where, {off(hold),on(auto),NOMinal} is:
off(or hold):
Sets the auto range to off.
on(or auto):
Sets the auto range to on.
NOMinal: Sets the range No.
Example
SEND> FUNC:RANG:AUTO AUTO<NL> //Sets to
auto range.
SEND> FUNC:RANG:AUTO off<NL> //Sets auto
range to off.
Query Syntax
FUNC:RANGe:AUTO?
Query Response {HOLD,AUTO,NOM}
Example
SEND> FUNC:RANG:AUTO?<NL>
RET>
auto<NL>
11.7.6 FUNCtion:MONitor1 /2
The FUNCtion:MONitor1 and FUNCtion:MONitor2 commands set the two
monitor parameter.
Command
Syntax
FUNC:MONitor1 {off, Z, D, Q, THR, THD, R, X, G, B,
Y, ABS, PER VAC, IAC}
FUNC:MONitor2 {off, Z, D, Q, THR, THD, R, X, G, B,
Y, ABS, PER VAC, IAC}
Parameter
Where, {off, Z, D, Q, THR, THD, R, X, G, B, Y, ABS,
PER VAC, IAC}
Example
SEND> FUNC:MON1 Z<NL>
Query Syntax
FUNC:MON1?
FUNC:MON2?
Query Response {off, Z, D, Q, THR, THD, R, X, G, B, Y, ABS, PER
VAC, IAC}
Example
SEND> FUNC:MON1?<NL>
RET>
off<NL>