LCR-6000 Series User Manual
75
6.1.7 Result
If the Result setting it set to Auto, the meter will
automatically send out the measurement results each time
a test is finished; this setting is convenient especially
when the meter is working with a sorting machine. The
meter will start a test after receiving the trigger signal and
then returns the test result to the sorting machine
without the need to receive a ‘fetch?’ command from
either the sorting machine or the control PC.
To set up the Result feature
Step 1. Press the [Measure] or [Setup] key.
Step 2. Press the [SYSTEM] bottom soft-key.
Step 3. Use the cursor key to select the [RESULT]
field.
Step 4. Use the soft-keys to set.
Soft-key
Function
FETCH
The meter will not send out the test
result after the test.
AUTO
The meter will send out the test result
after the test.
6.1.8 DATA BUFFER
The data buffer setting sets the maximum number of
record counts that the internal buffers can store; Settable
range 1~10000.
To set up the Data Buffer feature
Step 1. Press the [Measure] or [Setup] key.
Step 2. Press the [SYSTEM] bottom soft-key.
Step 3. Use the cursor key to select the
[DATA BUFFER] field.
Step 4. Enter the number of sets using the entry keys
or use the soft-key to select the maximum
number of sets.
Soft-key
Function
MAX
Sets the maximum record counts to be
10000 directly.