ASR-2000 Series User Manual
192
F
AQ
•
The accuracy does not match the specification.
•
How frequently should the power source be calibrated?
•
Is it proper to combine 2 or 3 units to reach 1P3W or 3P4W output?
The accuracy does not match the specification.
Make sure the device is powered On for at least 30 minutes, within
+18°C~+28°C. This is necessary to stabilize the unit to match the
specification.
How frequently should the power source be calibrated?
The ASR-2000 should be calibrated by an authorized service center
at least every 2 years.
For details regarding calibration, contact your local dealer or
GWInstek.
Is it proper to combine 2 or 3 units to reach the 1P3W or
3P4W output?
Not available. ASR-2000 doesn’t support 1P3W or 3P4W output
function. Only support 1P2W output type.
For more information, contact your local dealer or GWInstek at
www.gwinstek.com / [email protected].