TEST MODE FUNCTION
159
Simulate Mode
Simulate Mode Overview
Background
The Simulate function, which works in AC+DC-INT
mode only, is used to test power supply fluctuation.
This function is able to simulate common
abnormalities in mains power such as fluctuations
in voltage, phase and frequency. These simulations
can be run as one-off anomalies or cyclic anomalies.
Setting Screen
Overview
Number of Repeat
Sync Code
Step Wave
Step Frequency
Step Voltage
Step
ON/OFF Phase
Step Time
Test Mode
V RANGE
Recall Test
Save Test
Run Test
Step Overview
The Simulate function is comprised of 6 steps. Each
step is run sequentially in the following order:
Initial, Normal1, Trans1, Abnormal, Trans2,
Normal2 and Initial.
Initial
The Initial step is used as the initial
and final settings of the waveform
simulation. This is the standby step
before the test starts and the standby
step after the test ends.
Normal1 This step configures the normal output
conditions that precede the abnormal
conditions.