GUIDO RAYOS X, S.A.
NESTOMAT 6050
41
MAN-012
Sept.00
Ed. 1 / Rev. 3
•
Set Simulator switch at "40/99".
•
Twist
P8
Potentiometer to read 99.0% at the Oxygen Concentration Display on
the Incubator Control.
•
Set Simulator switch at "18/21".
•
Twist
P9
Potentiometer to read 21.0% at the Oxygen Concentration Display on
the Incubator Control.
Because adjustments are mutual interactive, therefore have to be repeated steps 3 and 6
until to get the correct value.
Once all adjustments are performed, proceed as follows:
•
Turn off the Radiant Warmer by pressing the I/O Switch located at the rear side
of the trolley.
•
Disconnet the jumper placed at J15 (point 1 - Air temperature).
N
OTE
: I
T IS VERY IMPORTANT NOT TO FORGET TO DISCONNECT THIS JUMPER
.
I
F YOU DO NOT DISCONNECT IT AND WHEN THE UNIT IS WORKING
,
IT IS DICON-
NECTED THE AIR TEMPERATURE SENSOR THE UNIT WILL SWITCH OFF
.
•
Release the Simulator.
•
Connect all sensors wires on their locations.
•
Turn on the Radiant Warmer by pressing the I/O Switch and the ON/OFF key.
•
Activate on the Activation/Deactivation option all those available sensors on the
Radiant Warmer (Paragraph II.6).
•
Turn off the Radiant Warmer and then Turn on again to make all the above
effective.