GUIDO RAYOS X, S.A.
NESTOMAT 6050
40
MAN-012
Sept.00
Ed. 1 / Rev. 3
Turn on the Radiant Warmer by pressing the I/O Switch located at the rear side of the
trolley.
Press the ON/OFF key.
Press three times the "PRE." key.
The LCD will display :
AT=00.0 PT=00.0
OX=00.0 RH=00.0
(AT: Air Temp., PT: Patient Temp., OX: Oxygen, RH: Non operative parameter)
A. Air Temperature
•
Place a jumper at J15.
•
Set Simulator switch at "0".
•
Set DVM probes between
GND-TP8
and
TP1
. Twist
P1
Potentiometer to read
between 2 to 4 mV.
•
Set Simulator switch at "40/99".
•
Twist
P2
Potentiometer to read 40.0 ºC at the Air Temperature Display on the
Incubator Control.
•
Set Simulator switch at "18/21".
•
Twist
P1
Potentiometer to read 18.0 ºC at the Air Temperature Display on the
Incubator Control.
Because adjustments are mutual interactive, therefore have to be repeated steps 3 and 6
until to get the correct value.
B. Patient Temperature
•
Set Simulator switch at "0".
•
Set DVM probes between
GND-TP8
and
TP2
. Twist
P5
Potentiometer to read
between 2 to 4 mV.
•
Set Simulator switch at "40/99".
•
Twist
P4
Potentiometer to read 40.0 ºC at the Air Temperature Display on the
Incubator Control.
•
Set Simulator switch at "18/21".
•
Twist
P5
Potentiometer to read 18.0 ºC at the Air Temperature Display on the
Incubator Control.
Because adjustments are mutual interactive, therefore have to be repeated steps 3 and 6
until to get the correct value.
C. Oxygen Concentration
•
Set Simulator switch at "0".
•
Set DVM probes between
GND-TP8
and
TP5
. Twist
P9
Potentiometer to read
5 mV maximum.
Содержание NESTOMAT 6050
Страница 31: ...GUIDO RAYOS X S A NESTOMAT 6050 31 MAN 012 Sept 00 Ed 1 Rev 3 1 2 2 3 4 5 6 7 8 9 10...
Страница 37: ...GUIDO RAYOS X S A NESTOMAT 6050 37 MAN 012 Sept 00 Ed 1 Rev 3...
Страница 38: ...GUIDO RAYOS X S A NESTOMAT 6050 38 MAN 012 Sept 00 Ed 1 Rev 3...
Страница 47: ...GUIDO RAYOS X S A NESTOMAT 6050 47 MAN 012 Sept 00 Ed 1 Rev 3...