supply tube.
Check that the Circuit Pressure gauge reads zero with the Gas inlet disconnected.
Occlude the samall aperture in the Positive End Expiratory Pressure (PEEP) cap
on the T-piece with your thumb.
Turn the Inspiratory Pressure (PIP) control clockwise until the manometer
needle stops. This may require few rotations due to the fine thread pitch of the
valve. The manometer now indicates the current Maximum Pressure Relief
setting.
5
To adjust the maximum Pressure Relief:
Determine the current Maximum Pressure relief setting as described in step 4 .
Occlude the PEEP cap aperture with your thumb.
Rotate the Maximum Pressure Relief control to the desired Maximum Pressure
Relief setting.
6
To adjust the Inspiratory Pressure (PIP) valve:
Check the test lung is attached to the T-piece and occlude the PEEP cap aperture
with your thumb.
Adjust the Inspiratory Pressure contol knob to the desired peal inspiratory
pressure as displayed on the manometer.
7
To adjust the Positive End Expiratory Pressure (PPEP) valve:
Remove your thumb form the PEED cap aperture to allow gas to flow through