GUIDO RAYOS X, S.A.
NESTOMAT 6050
16
MAN-012
Sept.00
Ed. 1 / Rev. 3
1=OXYG, 2=HUMID
NOTE:
D
UE TO THE WEAR OF THE
O
XYGEN CELL SENSOR IT
IS
RECOMMENDED TO PERFORM PERIODICALLY ITS CALIBRATION PROCEDURES
.
Check the Oxygen Sensor is pluged on the front panel and its housing is
not located inside the Tent Hood, allow to stabilize.
Press
1
, the LCD will display,
AIR=__._ ºC
OX: 1=PRG, 2=CAL
Press
2
, the LCD will display ,
OX CALIBR (21 %)
PRESS ENTER
By pressing
ENTER
, the Oxygen Sensor is already calibrated with a new
reference.
New screen displayed ;
AIR=__._ ºC
OX: 1=PRG, 2=CAL
II.6. ALARMS
The NESTOMAT 6050 Alarm System informs to the medical attendant of any
incidence may occur. The alarm is visible and audible.
NOMENCLATURE OF SYMBOLS IN ALARM
MESSAGES
+ Excess temperature alarm
-
Failing temperature alarm
= Temperautre O.K.
*
Desactive sensor
A) AIR CONTROL MODE
Содержание NESTOMAT 6050
Страница 31: ...GUIDO RAYOS X S A NESTOMAT 6050 31 MAN 012 Sept 00 Ed 1 Rev 3 1 2 2 3 4 5 6 7 8 9 10...
Страница 37: ...GUIDO RAYOS X S A NESTOMAT 6050 37 MAN 012 Sept 00 Ed 1 Rev 3...
Страница 38: ...GUIDO RAYOS X S A NESTOMAT 6050 38 MAN 012 Sept 00 Ed 1 Rev 3...
Страница 47: ...GUIDO RAYOS X S A NESTOMAT 6050 47 MAN 012 Sept 00 Ed 1 Rev 3...