background image

T02.0.3X.6C-07 

Mounting and operating manual OXY 3610 MP

  

Page 9 von 14 

 

7.1.2 

Average 

ambient absolute pressure “

PABS

 

Parameter 

button 1 

Value 

 button 2 / 3 

Description 

 

PABS 

500  2000

 

Necessary for a correct evaluation of the state of the electrode. If the 
application works for example at an altitude of 350 m above sea level, 
980mbar is a correct setting. Please refer to chapter 10 Ambient pres-
sure as a function of the altitude above sea level.

 

7.1.3 

Display damping “

F

i

LT

“ 

Parameter 

button 1 

Value 

 button 2 / 3 

Description 

F

i

LT

 

0  15.0

 

By this factor the ‘speed’ of the display can be adopted to the application. This 
is intended to reduce instabilities in the measured value as they appear in 
some specialised applications. 

Enter the desired value by using keys 2 (down) and 3 (up). Input range: 0...15 
(0: fast display, 15: slow display. Setting ex works: 10. 

7.1.4 

Save the last parameter values 

Parameter 

button 1 

Value 

 button 2 / 3 

Description 

8888

 

8888

 

The values will be saved. The device will start again (The display is shown 
8888). 

 

Bring back the jumper j1 to the ‚parking position after configuration (one contact connected, the 
second contact on air)!  
Then the device can be calibrated by key 1 and the configuration data is protected. 

Error and System Messages 

Display

 

Description 

Possible fault cause 

Remedy 

FE  1 

measuring range exceeded 

Calibration is wrong 

Recalibrate the transducer. 

FE  2 

Measuring values below 
measuring range 

Wrong signal 

Check the connections, cable and 
electrode. 

FE  7 

System fault 

Error in device 

Disconnect from supply and reconnect. 
If error remains: return to manufacturer 

FE  9

 

Input signal is not valid 

Sensor not connected or  
cable defective 
Temperature out of range 

Check electrode, cable and connec-
tions 
Check temperature 

8.8.8.8 

Segment test 

The transducer performs a display test for 2 seconds after power up. Af-
ter that it will change to the display of the measuring. 

Input signal is not valid 

Sensor not connected 

Check sensor, cable and connections 

Permissible input range is ex-
ceeded 

Check if not a wrong sensor is con-
nected. 
Replace sensor. 

Содержание OXY 3610 MP

Страница 1: ...3X 6C 07 Operating Manual Dissolved Oxygen Measuring transducer OXY 3610 MP Keep for later use www ghm group de Companies Brands of the GHM Englisch...

Страница 2: ...OF THE OXYGEN MEASURING TRANSDUCER 8 7 CONFIGURATION OF THE DEVICE 8 7 1 CONFIGURATION PROCEDURE 8 7 1 1 OFFSET OFFS 8 7 1 2 Average ambient absolute pressure PABS 9 7 1 3 Display damping FiLT 9 7 1 4...

Страница 3: ...asured values This manual can be a valuable contribution All instructions have to be well understood and complied with To be sure that there is no risk arising due to misinter pretation of measured va...

Страница 4: ...h large amounts of water for several minutes To clothing remove contaminated clothing To eyes Flush with large amounts of water for at least 15 minutes obtain medical treatment After swallowing Give l...

Страница 5: ...repair or maintenance With costs The conceptual design particularly when connecting with other devices has to be done thor oughly Internal connections of other devices e g from ground with protective...

Страница 6: ...sealing cable fitting b shim c and sealing ring d Pull out the connection socket f Thread cable through parts b c d and into connector housing e To mount the connection cable 2 3 or 4 wire depending...

Страница 7: ...normal operation the oxygen content is displayed in the unit mg l By pressing the Key 2 down the temper ature at the electrode can be shown pressing key 3 up shows the state of the electrode in The st...

Страница 8: ...k Calibration surrounding Oxygen electrodes are sensitive components Please read the referring manual prior to use During calibration the rating of the electrode is evaluated 100 means perfect state 4...

Страница 9: ...8888 The values will be saved The device will start again The display is shown 8888 Bring back the jumper j1 to the parking position after configuration one contact connected the second contact on air...

Страница 10: ...uction of the sensor GWO 3600 11 1 1 General The oxygen electrode is an active electrode consisting of a silver cathode and a lead anode using potassium hydrox ide KOH as an electrolyte In case of oxy...

Страница 11: ...ements This may also be the reason for errors in the calibration The membrane head is a spare part and can be ordered individually o Filling hole If the electrode is used at high temperatures or if it...

Страница 12: ...d vertically from above till the thread catches Then take up head with paper towel and screw onto electrode from underneath Electrolyte will be displaced from the membrane head flowing over Mop up exc...

Страница 13: ...2 years at nominal condition Display 1 5 of measured value 0 2 mg l Output signal 0 2 FS Connection 4 20 mA Zweileiter Spannung Dreileiter bzw Vierleiter Add output signal 0 2 FS Auxiliary energy sup...

Страница 14: ...x 80 x 55 mm without elbow type plug and sensor sockets Mounting With fixing holes for wall mounting in housing accessible after cover has been removed Mounting distance 50 x 70mm max shaft diameter...

Отзывы: