7.1.2
Average
ambient absolute pressure “
PABS
“
Parameter
button 1
Value
button 2 / 3
Description
PABS
500 2000
Necessary for a correct evaluation of the state of the electrode. If the
application works for example at an altitude of 350 m above sea level,
980mbar is a correct setting. Please refer to chapter 10 Ambient pres-
sure as a function of the altitude above sea level.
7.1.3
Display damping “
F
i
LT
“
Parameter
button 1
Value
button 2 / 3
Description
F
i
LT
0 15.0
By this factor the ‘speed’ of the display can be adopted to the application. This
is intended to reduce instabilities in the measured value as they appear in
some specialised applications.
Enter the desired value by using keys 2 (down) and 3 (up). Input range: 0...15
(0: fast display, 15: slow display. Setting ex works: 10.
7.1.4
Save the last parameter values
Parameter
button 1
Value
button 2 / 3
Description
8888
8888
The values will be saved. The device will start again (The display is shown
8888).
Bring back the jumper j1 to the ‚parking position after configuration (one contact connected, the
second contact on air)!
Then the device can be calibrated by key 1 and the configuration data is protected.
8
Error and System Messages
Display
Description
Possible fault cause
Remedy
FE 1
measuring range exceeded
Calibration is wrong
Recalibrate the transducer.
FE 2
Measuring values below
measuring range
Wrong signal
Check the connections, cable and
electrode.
FE 7
System fault
Error in device
Disconnect from supply and reconnect.
If error remains: return to manufacturer
FE 9
Input signal is not valid
Sensor not connected or
cable defective
Temperature out of range
Check electrode, cable and connec-
tions
Check temperature
8.8.8.8
Segment test
The transducer performs a display test for 2 seconds after power up. Af-
ter that it will change to the display of the measuring.
Input signal is not valid
Sensor not connected
Check sensor, cable and connections
Permissible input range is ex-
ceeded
Check if not a wrong sensor is con-
nected.
Replace sensor.