background image

T02.0.3X.6C-07 

Mounting and operating manual OXY 3610 MP

  

Page 11 von 14 

 

11.1.2  Design 

 

 

 

The electrode housing is made of ABS. With the exception of the electrode shaft all parts need to be maintained 
regularly and be replaced if necessary. 

o Protective flask

: The protective flask is used to moisten the membrane. This prolongs service life of the electrode. 

The protective flask contains water. Attention! Use water only; never use potassium chloride (KCI); this is only required 
for storage of pH-electrode. 

o Diaphragm head

: The membrane head is covered with a Teflon membrane. It will be filled with KOH electrolyte and 

screwed onto the electrode shaft (no air bubbles). Damages in the membrane, large air bubbles or air bubble rings in 
the membrane head will result in erroneous measurements. This may also be the reason for errors in the calibration. 
The membrane head is a spare part and can be ordered individually. 

o Filling hole

: If the electrode is used at high temperatures or if it has been stored without its protective flask for a 

longer period of time, some electrolyte will be lost due to evaporation. Please refer to Refilling description below. 

11.2  When is electrode maintenance required 

 

Attention when working with electrolyte! 
The electrolyte is caustic! (strong base, KOH) 
Avoid contact with skin and eyes. 
Protective equipment is necessary! 
Irritates skin and mucous membranes, caustic. 

In the case of contact with the eyes, immediately clean with clear water about 15 minutes

 

 

Eye protective is necessary!

 

 

Protective gloves are necesarry when filling the electrode! Do not touch the electrolyte with your 
bare hands. When contact with the skin immediately clean with water at least for 15 minutes. 

 

 

Electrode shaf 

Diaphragm 

Replenishing hole 

Protective flask 

Содержание OXY 3610 MP

Страница 1: ...3X 6C 07 Operating Manual Dissolved Oxygen Measuring transducer OXY 3610 MP Keep for later use www ghm group de Companies Brands of the GHM Englisch...

Страница 2: ...OF THE OXYGEN MEASURING TRANSDUCER 8 7 CONFIGURATION OF THE DEVICE 8 7 1 CONFIGURATION PROCEDURE 8 7 1 1 OFFSET OFFS 8 7 1 2 Average ambient absolute pressure PABS 9 7 1 3 Display damping FiLT 9 7 1 4...

Страница 3: ...asured values This manual can be a valuable contribution All instructions have to be well understood and complied with To be sure that there is no risk arising due to misinter pretation of measured va...

Страница 4: ...h large amounts of water for several minutes To clothing remove contaminated clothing To eyes Flush with large amounts of water for at least 15 minutes obtain medical treatment After swallowing Give l...

Страница 5: ...repair or maintenance With costs The conceptual design particularly when connecting with other devices has to be done thor oughly Internal connections of other devices e g from ground with protective...

Страница 6: ...sealing cable fitting b shim c and sealing ring d Pull out the connection socket f Thread cable through parts b c d and into connector housing e To mount the connection cable 2 3 or 4 wire depending...

Страница 7: ...normal operation the oxygen content is displayed in the unit mg l By pressing the Key 2 down the temper ature at the electrode can be shown pressing key 3 up shows the state of the electrode in The st...

Страница 8: ...k Calibration surrounding Oxygen electrodes are sensitive components Please read the referring manual prior to use During calibration the rating of the electrode is evaluated 100 means perfect state 4...

Страница 9: ...8888 The values will be saved The device will start again The display is shown 8888 Bring back the jumper j1 to the parking position after configuration one contact connected the second contact on air...

Страница 10: ...uction of the sensor GWO 3600 11 1 1 General The oxygen electrode is an active electrode consisting of a silver cathode and a lead anode using potassium hydrox ide KOH as an electrolyte In case of oxy...

Страница 11: ...ements This may also be the reason for errors in the calibration The membrane head is a spare part and can be ordered individually o Filling hole If the electrode is used at high temperatures or if it...

Страница 12: ...d vertically from above till the thread catches Then take up head with paper towel and screw onto electrode from underneath Electrolyte will be displaced from the membrane head flowing over Mop up exc...

Страница 13: ...2 years at nominal condition Display 1 5 of measured value 0 2 mg l Output signal 0 2 FS Connection 4 20 mA Zweileiter Spannung Dreileiter bzw Vierleiter Add output signal 0 2 FS Auxiliary energy sup...

Страница 14: ...x 80 x 55 mm without elbow type plug and sensor sockets Mounting With fixing holes for wall mounting in housing accessible after cover has been removed Mounting distance 50 x 70mm max shaft diameter...

Отзывы: