T02.0.3X.6C-07 

Mounting and operating manual OXY 3610 MP

  

Page 13 von 14 

 

12 

Decommissioning, Reshipment and disposal  

12.1  Decommissioning 

Always disconnect the device before from its supply before decommission e.g. at fuse. Valid general safety require-
ments shall be observed. 
Please also make sure that connected Loads are disconnected also and are in a safe state.

 

12.2  Reshipment and disposal 

 

Alle Geräte, die an den Hersteller zurückgeliefert werden, müssen frei von Gefahrstof-
fen sein. Gefahrstoffe am Gehäuse können Personen oder die Umwelt gefährden. 

 

Verwenden Sie zur Rücksendung des Geräts, insbesondere wenn es sich um ein noch funktio-
nierendes Gerät handelt, eine geeignete Transportverpackung. 
Legen Sie dem Gerät das ausgefüllte Rücksendeformular der GHM-Homepage unter 
http://www.ghm-messtechnik.de/downloads/ghm-formulare.html bei. 

 

Das Gerät darf nicht über die Restmülltonne entsorgt werden.  
Senden sie das Gerät an uns zurück (ausreichend frankiert, siehe auch Hinweise oben). 
Wir entsorgen das Gerät sachgerecht und umweltschonend. 

13 

Specification 

Measuring range 

0.00 

… 25.00 mg/l dissolved oxygen

 

Output signal 

See 2.2 

Fehler! Verweisquelle konnte nicht gefunden werden. 

Accuracy

:

 (at nominal temperature)

 

Approx. 2 years at nominal condition

 

Display 

±1.5% of measured value ±0.2 mg/l

 

Output signal 

±0.2 % FS

 

Connection 

4 - 20 mA  (Zweileiter) 
Spannung  (Dreileiter bzw. Vierleiter)

 

Add. output signal 

±0.2 % FS

 

Auxiliary energy:

 

(supply voltage)

 

Uv = 12 - 30 V DC (4-20mA) 
Uv = 18 - 30 V DC (0-10V)   
or see 2.2 

Fehler! Verweisquelle konnte nicht gefunden werden. 

Reverse voltage protection 

50V permanent

 

Perm. impedance 

(at 4-20mA)

 

R

A

(Ohm) < (Uv - 12V) / 0.02A ) 

Example: for Uv = 18V 
R

A

 < (18V - 12V) / 0.02A     =>   R

A

 < 300 Ohm 

Permissible load 

(at 0-...V)

 

R

L

(Ohm) > 3000 Ohm

 

Electrode 

Special oxygen electrode GWO3600MU (in scope of supply!) 

Electrode type 

active diaphragm type, with integrated NTC resistor 

Response time 

95% in 10 sec., depending on temperature 

Operation life 

3 years or more, depending on proper maintenance 

Operating pressure 

max. 3 bar 

Mounting diameter 

Ø 12,0 ±0,2 mm (also suitable for ½" glanding) 

Length 

approx. 220 mm (incl. anti-buckling glanding), Mounting length: approx. 
110 mm 

Elektrodenanschluss 

5 polige verschraubbare Diodenbuchse

 

Temperature compensation 

automatic, 0 ... 50 °C 

Electrode 

Special oxygen electrode GWO3600MU (in scope of supply!) 

Electrode type 

active diaphragm type, with integrated NTC resistor 

Response time 

95% in 10 sec., depending on temperature 

Operation life 

3 years or more, depending on proper maintenance 

Operating pressure 

max. 3 bar 

Mounting diameter 

Ø 12,0 ±0,2 mm (also suitable for ½" glanding) 

Length 

approx. 220 mm (incl. anti-buckling glanding), Mounting length: approx. 
110 mm 

Содержание OXY 3610 MP

Страница 1: ...3X 6C 07 Operating Manual Dissolved Oxygen Measuring transducer OXY 3610 MP Keep for later use www ghm group de Companies Brands of the GHM Englisch...

Страница 2: ...OF THE OXYGEN MEASURING TRANSDUCER 8 7 CONFIGURATION OF THE DEVICE 8 7 1 CONFIGURATION PROCEDURE 8 7 1 1 OFFSET OFFS 8 7 1 2 Average ambient absolute pressure PABS 9 7 1 3 Display damping FiLT 9 7 1 4...

Страница 3: ...asured values This manual can be a valuable contribution All instructions have to be well understood and complied with To be sure that there is no risk arising due to misinter pretation of measured va...

Страница 4: ...h large amounts of water for several minutes To clothing remove contaminated clothing To eyes Flush with large amounts of water for at least 15 minutes obtain medical treatment After swallowing Give l...

Страница 5: ...repair or maintenance With costs The conceptual design particularly when connecting with other devices has to be done thor oughly Internal connections of other devices e g from ground with protective...

Страница 6: ...sealing cable fitting b shim c and sealing ring d Pull out the connection socket f Thread cable through parts b c d and into connector housing e To mount the connection cable 2 3 or 4 wire depending...

Страница 7: ...normal operation the oxygen content is displayed in the unit mg l By pressing the Key 2 down the temper ature at the electrode can be shown pressing key 3 up shows the state of the electrode in The st...

Страница 8: ...k Calibration surrounding Oxygen electrodes are sensitive components Please read the referring manual prior to use During calibration the rating of the electrode is evaluated 100 means perfect state 4...

Страница 9: ...8888 The values will be saved The device will start again The display is shown 8888 Bring back the jumper j1 to the parking position after configuration one contact connected the second contact on air...

Страница 10: ...uction of the sensor GWO 3600 11 1 1 General The oxygen electrode is an active electrode consisting of a silver cathode and a lead anode using potassium hydrox ide KOH as an electrolyte In case of oxy...

Страница 11: ...ements This may also be the reason for errors in the calibration The membrane head is a spare part and can be ordered individually o Filling hole If the electrode is used at high temperatures or if it...

Страница 12: ...d vertically from above till the thread catches Then take up head with paper towel and screw onto electrode from underneath Electrolyte will be displaced from the membrane head flowing over Mop up exc...

Страница 13: ...2 years at nominal condition Display 1 5 of measured value 0 2 mg l Output signal 0 2 FS Connection 4 20 mA Zweileiter Spannung Dreileiter bzw Vierleiter Add output signal 0 2 FS Auxiliary energy sup...

Страница 14: ...x 80 x 55 mm without elbow type plug and sensor sockets Mounting With fixing holes for wall mounting in housing accessible after cover has been removed Mounting distance 50 x 70mm max shaft diameter...