c) Clear manual recordings:
If data sets have been stored, they can be deleted with the “store” key:
Press “store” for 2 seconds
: Call menu “Clear”
Select with:
or
.
Clear nothing (cancel)
Clear all data sets
Clear the latest data set
Confirm selection and quit menu “Clear”
8.2 Automatic recording with selectable cycle time “Func CYCL”
If logger function “Func CYCL” is selected (see “Configuration”) the device will automatically record
measuring values at intervals of the set cycle time.
The logger’s cycle time can be set from 1s to 60min (see “Configuration”).
Up to 10000 measurements can be saved if logger function “Func CYCL” is selected.
a) Start recording:
Press “store” for 2 seconds:
Start menu, press
again: automatic recording is started
Each storage process is signalled by shortly displaying “St.XXXX”, where XXXX is the number of the
saved data set.
If the logger memory is full, the recording stops automatically and the display shows
b)
Stop recording:
Press “store” for 2 seconds:
If recording is running the “stop” menu is displayed
Select with
or
.
Do not stop recording (cancel)
Stop recording
If recording is running the “stop” menu is displayed
If you try to switch off the device while cyclic recording is active you will be asked
whether the recording should really be stopped.
The device can only be switched off if the recording is stopped. Auto-off function is
deactivated as long as cyclic recording is active.
c) Clear recordings:
Press “store” for 2 seconds:
If there are data sets stored and recording is already stopped the menu “Clear” is displayed
Select with
or
.
Do not stop recording (cancel)
Clear all data sets
Clear latest data set
Confirm selection and quit menu “Clear”