background image

H72.0.11.6C-04 

 Operating Manual GMH 3451 

page 6 of 18 

 

 _____________________________________________________ 

_____________________________________________________________________________ 

 

4.4 Pop-up clip 

Handling: 

  Pull at label “open” in order to swing open the pop-up clip. 

  Pull at label “open” again to swing open the pop-up clip further. 

 

Pop-up clip closed 

 

Pop-up clip at position 90° 

 

Pop-up clip at position 180° 

Function: 

  The device with a closed pop-up clip can be plainly laid onto a table or attached to a belt, etc. 

  The device with pop-up clip at position 90° can be set up on a table, etc. 

  The device with pop-up clip at position 180° can be suspended from a screw or the magnetic holder 

GMH 1300. 

 

Device

 

attached

 

to

 

a

 

belt

 

 

 

 

Device

 

set

 

up

 

on

 

a

 

table

 

 

Device

 

suspended

 

from

 

magnetic

 

holder

 

GMH

 

1300

 

5 Start 

Operation 

Turn device on via 

 key. 

After segment test 

 

 

the device displays some information on its configuration: 

  if cell correction scale was changed (cell correction scale unequal 1.000) 

  

(see chapter 11 Automatic adjustment/calibration of cell correction) 

   if zero point or slope correction is active 

 

(see chapter 10 Adjustment of temperature input) 

After that the device is ready for measuring. 

Содержание GMH 3451

Страница 1: ...49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Please carefully read these instructions before use Please consider the safety instructions Please keep for future reference as of version 1 3 GMH...

Страница 2: ...6 ELECTRODES MEASURING CELLS 8 6 6 1 Design 8 6 6 2 Calibration Adjustment of measuring cells 8 6 7 TEMPERATURE COMPENSATION 8 6 7 1 Temperature compensation nLF according to EN 27888 8 6 7 2 Linear t...

Страница 3: ...ficant damage to property at non observance Attention This symbol warns of possible dangers or dangerous situations which can provoke damage to the device or environment at non observance Note This sy...

Страница 4: ...ion If and bAt is shown in the lower display the battery has been used up and needs to be replaced However the device will operate correctly for a certain time If bAt is shown in the upper display the...

Страница 5: ...shortly change over between measuring units only if InP SEt is selected press for 2 sec menu invoke configuration menu min max when taking measurements press shortly min or max value is displayed pre...

Страница 6: ...tached to a belt etc The device with pop up clip at position 90 can be set up on a table etc The device with pop up clip at position 180 can be suspended from a screw or the magnetic holder GMH 1300 D...

Страница 7: ...d solids measurement the filtrate dry residue is determined by means of the conductivity and a conversion factor C tdS Well suited for easy concentration measurements of e g salt solutions The determi...

Страница 8: ...t see chapter 15 Accuracy Check Adjustment Service 6 7 Temperature compensation The conductivity of aqueous solutions depends on its temperature The temperature dependency is strongly dependent on the...

Страница 9: ...Input Selection of measured variable Cond Conductivity rESi Resistivity tdS Total dissolved solids SAL Salinity SEt Change over measured variables by Set key TDS measurement conversion factor only if...

Страница 10: ...uld correspond to output 0 V e g for 0 0 S cm 0 0 S cm 400 mS cm Measuring value that should correspond to output 1 V e g for 100 0 mS cm Set Corr Measurement correction Zero point adjustment offset o...

Страница 11: ...be used The software also allows easy configuration and starting of the logger When the logger is activated Func Stor or Func CYCL the hold function is no more available the key store is solely used...

Страница 12: ...rding Press store for 2 seconds Start menu press again automatic recording is started Each storage process is signalled by shortly displaying St XXXX where XXXX is the number of the saved data set If...

Страница 13: ...es In case you want to develop your own software we offer a GMH3000 development package including A universally applicable Windows functions library GMH3000 DLL with documentation that can be used by...

Страница 14: ...re different from default settings the device will shortly display Corr after turned on Default setting for offset and scale are off 0 0 i e inputs are not changed Zero point correction Displayed valu...

Страница 15: ...orrect pH calibration The device provides the following functions to help with this 12 1 Calibration interval C Int You can input the interval after which the device reminds you to recalibrate The int...

Страница 16: ...rer for repair Err 1 Measured value above allowable range Check pressure not within sensor range measuring value to high Sensor defective Return to manufacturer for repair Err 2 Measured value below a...

Страница 17: ...at housing or sensor may be a risk for persons or environment Use an adequate transport package for reshipment especially for fully functional devices Please make sure that the device is protected in...

Страница 18: ...l symbols Additional functions Min max hold Adjustment Calibration Cell correction manually or automatically via selectable reference solution Data logger Real time clock Cyclic 10000 data sets cycle...

Отзывы: