H72.0.11.6C-04
Operating Manual GMH 3451
page 11 of 18
_____________________________________________________
_____________________________________________________________________________
(*)
If logger memory contains data sets parameters marked with (*) cannot be called. You have to clear memory to
change these parameters!
(**)
If logger is running parameters marked with (**) cannot be called.
8 Data
Logger
No logger operation possible with auto-range! The measuring range has to be selected
explicitly – see chapter 7 “Configuration“ -
The device supports two different logger functions:
“Func-Stor”:
Manual recording by key press “store”
Additional input of measuring point (L-Id) is needed
“Func-CYCL”:
Automatic recording at intervals of set cycle time
The logger stores 2 measuring values per data set.
One data set consists of:
meas. value cond/ rES/ TDS/ SAL (one of them)
meas.
value
temperature
measuring
point
L-Id
(only
for
“Func-Stor”)
time and date (when data set is saved)
For the evaluation of the data the software GSOFT3050 (version V3.0 or higher) has to be used. The
software also allows easy configuration and starting of the logger.
When the logger is activated (Func Stor or Func CYCL) the hold function is no more available, the key
“store” is solely used for the operation of the logger functions.
8.1
Manual recording (“Func-Stor”)
a) Save measurements manually:
Up to 1000 measurements can be saved if logger function “Func store” is selected. (see “Configuration”):
Press “Store” shortly:
data set is saved (“St. XX” is displayed shortly,
where XX is the number of the data set)
Input of the measuring point
“L-Id”: Selection of measuring point via keys
or
.
Number 0…19999 or text assigned to number 1…40
(comfortable assignment of texts can be done with gratis software GMHKonfig)
Confirm input with
.
If logger storage is full, the following is displayed:
b) Read manual recordings:
Saved data sets can be viewed both with PC-software GSOFT3050 and directly on the device display.
Press “menu” for 2 seconds:
is displayed
“rEAd LoGG” is only displayed if data sets have been already stored! Otherwise the
configuration menu is displayed:
Press shortly:
Change between measuring values, measuring point and date+time
of the currently selected data set
or
Change between different data sets
Quit display of recordings