GREISINGER GMH 3451 Скачать руководство пользователя страница 13

H72.0.11.6C-04 

 Operating Manual GMH 3451 

page 13 of 18 

 

 _____________________________________________________ 

_____________________________________________________________________________ 

 

9 Universal 

output 

The output can be used as serial interface (for USB 3100, USB 3100 N, GRS 3100 or GRS 3105 
interface adapters) or as analogue output (0-1V). 

If none of both is needed, we suggest to switch the output off, because battery life then is extended. 

When operating with external supply or connected interface and measurement at solutions 
with earth connection, there may appear distortions or deviations of the measuring. In case of 
doubt disconnect supply/interface. 
The same is valid for the analogue output: Depending how the output is connected (e.g. even 
without isolation), in case of doubt do not measure in solutions with earth contact. 

9.1 Serial Interface 

By means of the serial interface and a suitable electrically isolated interface adapter (USB 3100, USB 3100 
N, GRS 3100 or GRS 3105) the device can be connected to a computer for data transfer.  
With the GRS 3105 up to 5 devices of the GMH3xxx- series can be connected to one interface (see also 
manual of GRS 3105). As a precondition the base addresses of all devices must not be identical, make sure 
to configure the base addresses accordingly (refer menu point “Adr.” in chapter 1 “Configuration“). 
To avoid transmission errors, there are several security checks implemented e.g. CRC. 

The following standard software packages are available: 

 

GSOFT3050

:  

Operation and read out of logger function for devices with integrated data logger 

 

GMHKonfig

Software for a comfortable editing of the device (e.g. Material selection…) 

 

EBS 20M / 60M

:  20-/60-channel software to display the measuring values 

In case you want to develop your own software we offer a 

GMH3000-development package

 including: 

  A universally applicable Windows functions library ('GMH3000.DLL') with documentation that can be 

used by the most programming languages. Suitable for Windows XP™, Windows Vista™, Windows 7™, 
Windows 8 / 8.1™, Windows 10™ 

  Programming examples Visual Studio 2010 (C#, C++ und VB), Testpoint™,LabView™ uvm.  

In addition to the operation at a PC the device can be operated with the 

GAM 3000 

device, to use the alarm 

function for simple supervision and controlling applications. Just connect a GAM 3000 to the interface, 
activate the alarm function of the GMH and the relays output is operating. 

The device has 2 channels: 

-  Channel 1: current measuring value Cond, rES, TDS oder SAL (base address) 
-  Channel 2: temperature value 

Supported functions:  

1  2  Code  Name/Function 

1 2 Code  Name/Function 

x x 0 

Read measurement value 

x x 200  Read min display range 

x x 3 

Read system state 

x x 201  Read max display range 

x    12 

Read ID number 

x x 202 

Read display range - unit 

x x 22 

Read min alarm rail (AL. - AL.Lo)  

x x 204 

Read display range - decimal point 

x x 23 

Read max alarm rail (AL. - AL.Hi)  

x   208 

Read # of channels 

x x 176  Read min measuring range 

x  222  Read power off time (Conf-P.oFF)  

x x 177  Read max measuring range 

x  223  Set power off time (Conf-P.oFF) 

x x 178  Read measuring range unit 

x  233  Read real time clock (CLOC)  

x x 179  Read measuring range decimal point 

x  234  Set real time clock (CLOC)  

x x 180  Read kind of measuring of sensor 

x  240  Reset 

x x 199  Read kind of measuring of display 

x  254  Program version 

 

The measuring-/ alarm- and display range values read back from the interface are 
always in the selected measurement unit! 
When using the interface, the auto-range-function should be turned off. If auto-range is 
activated, the returned value ar based on the resolution of the smallest ranges, there 
may be returned extreme values like 123400.0 µS/cm instead 123.4 mS/cm. 

 

 

 

Содержание GMH 3451

Страница 1: ...49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Please carefully read these instructions before use Please consider the safety instructions Please keep for future reference as of version 1 3 GMH...

Страница 2: ...6 ELECTRODES MEASURING CELLS 8 6 6 1 Design 8 6 6 2 Calibration Adjustment of measuring cells 8 6 7 TEMPERATURE COMPENSATION 8 6 7 1 Temperature compensation nLF according to EN 27888 8 6 7 2 Linear t...

Страница 3: ...ficant damage to property at non observance Attention This symbol warns of possible dangers or dangerous situations which can provoke damage to the device or environment at non observance Note This sy...

Страница 4: ...ion If and bAt is shown in the lower display the battery has been used up and needs to be replaced However the device will operate correctly for a certain time If bAt is shown in the upper display the...

Страница 5: ...shortly change over between measuring units only if InP SEt is selected press for 2 sec menu invoke configuration menu min max when taking measurements press shortly min or max value is displayed pre...

Страница 6: ...tached to a belt etc The device with pop up clip at position 90 can be set up on a table etc The device with pop up clip at position 180 can be suspended from a screw or the magnetic holder GMH 1300 D...

Страница 7: ...d solids measurement the filtrate dry residue is determined by means of the conductivity and a conversion factor C tdS Well suited for easy concentration measurements of e g salt solutions The determi...

Страница 8: ...t see chapter 15 Accuracy Check Adjustment Service 6 7 Temperature compensation The conductivity of aqueous solutions depends on its temperature The temperature dependency is strongly dependent on the...

Страница 9: ...Input Selection of measured variable Cond Conductivity rESi Resistivity tdS Total dissolved solids SAL Salinity SEt Change over measured variables by Set key TDS measurement conversion factor only if...

Страница 10: ...uld correspond to output 0 V e g for 0 0 S cm 0 0 S cm 400 mS cm Measuring value that should correspond to output 1 V e g for 100 0 mS cm Set Corr Measurement correction Zero point adjustment offset o...

Страница 11: ...be used The software also allows easy configuration and starting of the logger When the logger is activated Func Stor or Func CYCL the hold function is no more available the key store is solely used...

Страница 12: ...rding Press store for 2 seconds Start menu press again automatic recording is started Each storage process is signalled by shortly displaying St XXXX where XXXX is the number of the saved data set If...

Страница 13: ...es In case you want to develop your own software we offer a GMH3000 development package including A universally applicable Windows functions library GMH3000 DLL with documentation that can be used by...

Страница 14: ...re different from default settings the device will shortly display Corr after turned on Default setting for offset and scale are off 0 0 i e inputs are not changed Zero point correction Displayed valu...

Страница 15: ...orrect pH calibration The device provides the following functions to help with this 12 1 Calibration interval C Int You can input the interval after which the device reminds you to recalibrate The int...

Страница 16: ...rer for repair Err 1 Measured value above allowable range Check pressure not within sensor range measuring value to high Sensor defective Return to manufacturer for repair Err 2 Measured value below a...

Страница 17: ...at housing or sensor may be a risk for persons or environment Use an adequate transport package for reshipment especially for fully functional devices Please make sure that the device is protected in...

Страница 18: ...l symbols Additional functions Min max hold Adjustment Calibration Cell correction manually or automatically via selectable reference solution Data logger Real time clock Cyclic 10000 data sets cycle...