G r a s e b y M e d i c a l L t d .
4 — 2 4
4
I s s u e A ( A p r i l 2 0 0 2 )
3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l
V o l u m e t r i c I n f u s i o n P u m p s
Pressure display - calibration mode
If the pump is operating on AC mains supply, the pressure
transducer on the front door assembly sensor may be calibrated, so
you see this parameter:
A
B
C D
The bottom line of this screen shows four fields, marked on the above
example as A, B, C and D. They are described in the table below:
Field Range
Comment
A
0 to 1919
Represents motor position. It shows unref if
unref
the motor cam position is unreferenced. It shows
or *
*
in the first position if a motor stall was detected.
B
0 to 1919
Represents the position immediately before the
last motor home position realignment. If there is
no significant motor slippage, expected display
values are between 0 and 40 or 1880 and 1919.
C
0 to 255
Raw analog-to-digital pressure reading from the
display board. This value is read and the Message
Display value updated at least twice per second.
This permits occlusion alarm calibration to a
known pressure by adjusting two potentiometers,
see
Occlusion alarm adjustment
on page 8-15..
D
1, 0 or
-
Represents the pressure plate position
switch, where:
1 = cassette-in-position
0 = no cassette present
- = communication error when switch
is read.
Pressure transducer calibration screen
The default rate is 25 mL/h. To adjust the rate:
Press
Rate
and enter a rate with the number keys. The run light
flashes whenever a rate can be entered using number keys.
To accept the entered rate:
Press
Run
. The rate appears in the rate display.
To toggle the motor direction:
Press
Primary
. If the motor is running backwards, the pumping
indicator shows a minus sign. If the motor is running (forward or
backward) the run light is continuously lit.
To stop the motor.
Press
Hold
. When the motor stops the hold light is lit.
Содержание 3000/500
Страница 1: ...0473 Technical Service Manual Model 3000 500 and Micro 3100 505 Volumetric Infusion Pump...
Страница 2: ......
Страница 14: ...I n t r o d u c t i o n Chapter 1...
Страница 18: ...S p e c i f i c a t i o n Chapter 2...
Страница 29: ...O v e r v i e w o f P u m p O p e r a t i o n I n i t i a l C h e c k s Chapter 3...
Страница 47: ...V o l u m e t r i c P u m p M e n u s Chapter 4...
Страница 79: ...M e c h a n i c a l S y s t e m s Chapter 5...
Страница 95: ...E l e c t r o n i c C i r c u i t s Chapter 6...
Страница 96: ...M a i n t e n a n c e P r o c e d u r e s T r o u b l e s h o o t i n g Chapter 7...
Страница 129: ...D i s a s s e m b l y R e a s s e m b l y C a l i b r a t i o n I n s t r u c t i o n s Chapter 8...
Страница 151: ...I l l u s t r a t e d P a r t s L i s t Chapter 9...
Страница 178: ...E l e c t r i c a l D i a g r a m s Chapter 10...
Страница 189: ...W a v e f o r m D i a g r a m s Chapter 11...
Страница 198: ...R e m o t e M o n i t o r i n g C o n t r o l o f t h e p u m p Chapter 12...
Страница 225: ...I n d e x...